FIG.8

"MP" input board

PT1

SIGNAL INPUT RANGE SELECTION

The "MP" input board  for this signal conditioner,
does not have jumpers because it allows only one
signal input (0 to 20 mV). To calibrate the signal
conditioner, follow the adjustment and calibration
procedure.
The user can field adjust the tare using the trim-
mer PT1, located on the "MP" input board. See
Fig. 8.
The signal input 0 to 20 mV will be the standard
range for all orders, unless specified otherwise.

ADJUSTMENT AND CALIBRATION PROCEDURE

To calibrate the signal conditioner, use a calibrator or Vdc source from 0 to 20 mVdc.

1. Open the signal conditioner case. (Fig. 2).
2. Check the "MA" module, if the selected power supply is correct. (Table 1).
3. Connect the calibrator to the signal conditioner terminals 5 - 6.
4. Connect a digital multimeter, to the signal output terminals to be used.
5. Power up the signal conditioner with the correct power supply.
6. Adjust the  calibrator until it generates  the low signal level (0 Vdc).
7. Turn the "ZERO" trimmer (P1), located on the "ME" input module, until the multimeter

shows the desired low signal output level. (For example: 0 Vdc).

8. Adjust the calibrator until it generates the high signal level.
9. Turn the "GAIN" trimmer (P2), located on the "ME" input module, until the multimeter

shows the desired high output level. (For example: 10 Vdc).

10. Repeat steps 6 to 9, until the two values are correct.

TARE ADJUST PROCEDURE

After the connections between the signal conditioner, load cell and external power supply are
made, connect a multimeter to the output terminals. Do not forget that the load cell must be
free of load.
If the multimeter does not read 0 Vdc or 4 mA, then adjust the trimmer PT1, located on the
"MP" input board.
Check that the high level of the output signal does not change. If it does adjust the GAIN
trimmer (P2) located on the "ME" input module ( see Fig. 7).

The signal conditioner provides two different analog output
signals, both proportional to the signal input.

Output in Current: 4 to 20 mA, terminals 9 - 10
Output in Voltage: 0 to 10 Vdc, terminals 11 - 12

Do not use both outputs simultaneously. Only one selection
can be made.
The side label shows which one is selected.

SIGNAL OUTPUT module MS

OUTPUTS / SALIDAS
        mA          Vdc

INPUT / ENTRADA

SU

PPL

Y / A

L

IM

EN

T

A

C

IO

N

3    4    5    6

9   10   11   12

7

1

+

Lo

H i 

+

-

-

FIG.5

CHANGING THE SIGNAL OUTPUT

All signal conditioners are delivered as a standard version, with the analog output selected as
4 to 20 mA, unless specified otherwise. To select a 0 to 10 V output, remove jumpers E and
F as shown in Figure 6. Other non-standard output voltage and current ranges may be obtained
by adding and/or replacing resistors given in Tables 2 and 3.

Output

          Value in 

Ω 

for :

in mA.

R18

R24

R25

0 to 5

-----

100

-----

0 to 10

-----

49.9

-----

1 to 5

100 K

124

-----

0 to 20

-----

-------

24.9

NON STANDARD OUTPUT
CURRENTS

TABLE 2

TABLE 3

NON STANDARD OUTPUT
VOLTAGES

Output

        Value in K

Ω 

for :

in V.

R29

R30

R31

R32

±10

49.9

-----

200

----

0 to 1

-----

-----

11

100

0 to 5

-----

-----

100

100

1 to 5

-----

100

66.5

100

E

F

R18

R29

R30

R24 R25 R31 R32

FIG.6

Replace or add the indicated
resistors with the values shown
in Tables 2 and 3  for the desired
output.

5

 8

"-----" means "Resistor must not be installed"

Jumpers E and F : Closed only if the output is 4 to 20 mA.

Summary of Contents for CCT-80 Series

Page 1: ...M2218 1201 http www omega com e mail info omega com User sGuide CCT 80 Strain Gage Signal Conditioners...

Page 2: ...France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 07056 3017 FAX 49 07056 8540 Toll Free in Germany 0130 11 21 66 e mail germany omega...

Page 3: ...Changing the Power Supply 4 Signal Output module MS 5 Changing the Signal Output 5 Signal Input module ME Overview 6 CCT 80 Electrical Features Specifications 7 Signal Input Range Selection 8 Adjustm...

Page 4: ...the case the following must be checked The front label blue color must be in its correct position with terminals 1 and 7 power supply separated from the other terminals The three internal modules must...

Page 5: ...e Transmitter to control a machine or process where injury to personnel or damage to equipment or process may occur as a result of failure of the Transmitter Power Fuse supply value 230 Vac 50 mA 115...

Page 6: ...4 to 20 mA unless specified otherwise To select a 0 to 10 V output remove jumpers E and F as shown in Figure 6 Other non standard output voltage and current ranges may be obtained by adding and or re...

Page 7: ...if closed Adjust the low range level of the output Offset negative fine AMPLIFIER GAIN Jumper B if closed Gain at maximum level Jumper C if closed Gain at medium level Jumpers B and C opened Gain at...

Page 8: ...SUMPTION 1 5 VA TEST VOLTAGE 4 kVeff 50 Hz 1m This signal conditioner provides connection to 4 or 6 wire strain gages load cells which provide a nominal of 2 mV V or 3 mV V output with 350 impedance S...

Page 9: ...als to be used 5 Power up the signal conditioner with the correct power supply 6 Adjust the calibrator until it generates the low signal level 0 Vdc 7 Turn the ZERO trimmer P1 located on the ME input...

Page 10: ...ENTACION TARE ADJUST When internal tare adjust on the ME input board is not used the signal conditioner terminals 3 and 4 must not be connected POWER SUPPLY SIGNALOUTPUT Current SIGNALOUTPUT Voltage O...

Page 11: ...Lo H i TECHNICAL DATA WEIGHT 270 g HOUSING BASE Polycarbonate RAL 7032 UL 94 V 1 light grey IP 40 TERMINAL HOUSING COVER AND BLIND PLUGS Polycarbonate UL 94 V 2 dark grey IP 20 WIRE CROSS SECTION 4 mm...

Page 12: ...NOTES 11...

Page 13: ...rwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by...

Page 14: ...ystems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity...

Reviews: