7-1
BB701
7
Section 7 - Specifications
Target Plate Temperature Range:
-18.8 to 150°C (+0 to 300°F).
Accuracy:
±0.8°C + 1 Digit (±1.4°F) [worst case]
Display Accuracy:
±0.3°C (±0.6°F) [over entire scale]
RTD Accuracy:
±0.375°C (±0.75°F) [worst case]
Stability:
± 0.1°F or less
Ambient Environmental Conditions
Temperature:
4.4 to 29.4°C (40 to 85°F)
[target plate reaches 0°F]
4.4 to 43.3°C (40 to 110°F)
[target plate reaches 11.8°F]
Humidity:
0 to 90% RH, non-condensing
Target Plate Size:
63.5mm (2.5")
Cavity Emissivity:
0.95
Internal Control Sensor:
Platinum RTD, 100 ohm, .00385
Calibration Reference Sensor:
Platinum RTD, 100 ohm, .00385
Warm-up Time:
5 min., from ambient to 149°C (300°F)
Cool-Down Time:
5 min., from ambient to -18°C (0°F)
Power Requirements by Model
BB701:
115 VAC~, 50/60 Hz, 175W
BB701-230VAC:
230 VAC~, 50/60 Hz, 175W
Size:
19.5 x 35.5 x 36.5 mm
(7.75" H x 14.128" W x 15.5" L)
Weight:
4.5 kg (28 lb)
Installation Category
II