FEATURES
1) Super large LCD simultaneously display of HCHO level,
Temp., Relative humidity, Calendar (Y/M/D) and Time
(clock).
2) Three (3) smiley icons indicate indoor air quality levels
(0/0.08/0.8ppm), easy to understand HCHO concentration.
3) Stable Electrochemical type sensor for HCHO detection.
4) High alarm threshold is selectable.
5) Manual 0ppm calibration.
6) Audible/Visible LED warning for high HCHO concentration
1
o
ppm
AIR
TEMP
Humidity
M
D
LCD DISPLAY
Symbols:
1)ppm HCHO unit
2) icon
: 0ppm ~ 0.08ppm
3) icon
: 0.08ppm ~ 0.8ppm
4) icon
: 0.8ppm
5)Air Temp.: Ambient Temperature
6)MD : Month / Date
7)HM: Hour / Minute
8)TWA: Time Weighted Average
(8 hours)
9)STEL: Short-Term Exposure Limit
(15 minutes weighted average )
10)Logging: When it is in the recording mode
11)VOC, NO2, WBGT, hpa, mbar ,EXT TEMP,lps%,cfm/p are not
available in these models
:
INTRODUCTION
Thank you for purchasing desktop HCHO meter.
It is used to measure HCHO concentration,air temperature
and relative humidity with visible and audible alarms.
This HCHO meter is an ideal instrument for indoor air quality
(IAQ) diagnosis and HVAC system performance verification.
MATERIAL SUPPLIED
(1) Meter
(2) Universal Adaptor
(3) Operation manual
(4) 4 pcs batteries.
(5) USB cable
(6) White Box
POWER SUPPLY
The meter is powered by Super heavy duty batteries AAA x 4 pcs
(Recommended) or an AC adaptor(9V/0.5 A output)
2
KEYPAD (CONTROLS)
1) MODE : HCHO mode,MAX,TWA,STEL
2) SEL./R : Edit alarm function
Logger mode: Long press to key start/off.
3) TRH.M. : Temp. Humidity mode. Setting/ Next
4) UNIT
: Select Temp. unit
5) POWER: Turn meter on/off, alarm on/off.
OPERATION
Warning: Remove batteries when not in use to prevent battery
leakage (Out of warranty) and power consumption.
(a) POWER ON/OFF
Put battery or plug in adaptor, meter turns on automatically .
LCD shows current HCHO, RH & Temperature, Date and Time.
Three(3) smiley icons indicate the indoor air quality level and
appear on the top of first layer display.
Note:
(1) When low battery, meter will restart and logger will stop if
user switch to adaptor.
(2) Please use properly power source, input voltage:
100~240 VAC,50-60Hz
Output voltage: DC 7.5~9.0V
Output current: 0.5A,or the meter will be damaged.
(3) Battery life approx. 120 hours when continuous used
(b) TAKING MEASUREMENT
The meter starts the measurements when powered
on and response time is less than 30 sec.
Do not hold the meter close to any other source of HCHO.
Note:
(c) HCHO MODE
MODE
In the HCHO Mode, press
button to switch HCHO
display. The top layer of display will cycle from HCHO,
Maximum, TWA, STEL.
3
Note 1:
1) MAXIMUM:
The unit automatically records maximum reading
and minimum reading after powered on.(MINIMUM is
always 0.00ppm for HCHO meter)
2) TWA: ( Time Weighted Average 8 hours)
The meter keeps updating the reading every minute. If
the meter has been powered on for less than 8 hours,
the TWA value will be the weighted average of readings
since powered on.
3)STEL: (Short-Term Exposure Limit in 15 minutes weighted
average)The meter keeps updating readings every
minute. If the meter has been powered on for less than
15 minutes, the STEL value shows the weighted average
of reading taken since powered on.
4)RESET MAX:
Press
button to get MAX reading and
MODE
switch TWA or STEL.Under MAX mode, Long
press
button to clear all the values.
MODE
ppm
MAX
ppm
TWA
STEL
ppm
ppm
(d) TRH MODE : Temperature / Relative Humidity mode
TRH
In the
Mode, the meter displays both
related humidity and air Temperature
simultaneously.
(e) Temperature unit (°C or °F) selection
Press
button to toggle the
UNIT
temperature unit (°C or°F).
o
AIR
TEMP
HUMIDITY
+ TEMPERATURE
HUMIDITY
o
AIR
TEMP
HUMIDITY
o
AIR
TEMP
HUMIDITY
4
MAINTENANCE
Cleaning and storage:
1). The meter should be cleaned with a damp cloth and mild
detergent when necessary.
2). Store the meter in an area with moderate temperature
and humidity.
MANUAL CALIBRATION TO ZERO
Before start HCHO manual calibration make sure the area is
ventilation and no HCHO environment. Ensure the calibration
completed, to use adaptor during calibration is recommended.
HCHO meter is completed calibration in factory before
packed. Manual calibration is only for meter inaccuracy.
When power on, long press
button more
UNIT
than 6 seconds, the value will turn into “00000"
and flash
seconds.
60
After calibration completed, display will show
full LCD icon for a second. Meter will restart
automatically to return normal measurement.
BATTERY INDICATION
When battery is low (HCHO consumes power ),the reading
of HCHO will be terribly high, could be couple thousand in
ppm. “LO BAT” is flashing at the bottom of the screen,
near calendar and time. Replace new battery when LO BAT
shows.
SETTING DATE & TIME
a) Real time and date:
The meter shows Mo./Date and Hr/Min on the 3rd layer of
display, and each cycle is 16 seconds.
b) Month/Date and Hour/Min
Press
button to enter the real date and
MODE+POWER
time setting. The time default is 24-hours format.
M
D
5
HCHO Hi alarm setting.
Enter HCHO alarm setting.
+
SEL/R POWER
- ->
- ->
- ->
- ->
Change digit.
UNIT
Number goes up.
SEL/R
Number goes down.
TRH/M
After set up
+
to save and exit setting.
SEL/R POWER
ALARM SETTING
Set up alarm function:
In normal measurement, press
buttons ,
SEL/R+POWER
bell will appear alarm function. Once set up, the red
LED will flash if the data exceeding the setting data.
SETTING DATE & TIME
The meter interchangely shows current date and time
on the 3rd layer of display, and each cycle is
6 seconds.
The time default is 24-hours format.
+
-->Enter date/time mode
MODE POWER
--> Number goes up.
SEL/R
--> Number goes down.
TRH/M
-->Edit Month/date or Hour/Minute.
MODE
-->Select to edit Month or date, Hour or Minute.
UNIT
After set up, press
+
to save and exit.
MODE POWER
HCHO alarm
on
HCHO alarm
off
o
ppm
AIR
TEMP
Humidity
M
D
o
ppm
AIR
TEMP
Humidity
M
D
HCHO alarm ON/OFF:
In normal function, short press
bell icon appear
POWER
alarm on, short press
again to turn off.
POWER
o
AIR
TEMP
Humidity
H M
TEMP. HUMIDITY OFFSET
Make sure the meter is off, press
button
SEL/R+TRH/M
then plug in the adaptor at the same time to enter offsetting
mode.
WARNING: DO NOT USE THIS FUNCTION WITHOUT
STANDARD INSTRUMENT.
Example: Standard instrument is 30.2°C,
meter shows 30.0°C.Enter temp. offset mode, press
till 0.2 appear.
TRH/M
Offset range of Temp.and RH as below:
Temp.
:-9.9°C to +9.9°C / -9.9°F to +9.9°F
RH%
:-9.9% to +9.9%
a) Humidity offset:
Increase value
Decrease value
b) Air temp. offset
Increase value
Decrease value
After setting up short press
to save setting and
,
POWER
“%”
icon will flash a second. Then replug adaptor to return normal
measurement.
Logger setup:
When using datalogger function, please use batteries and adaptor
at the same time to avoid power outages and data loss.
A. Install logger driver for Windows
The PL2303_Prolific_DriverInstaller_v110.exedriver is for
Windows operation system, follow the installation steps
hereunder:
1.Put the CD into CD driver, click
PL2303_Prolific_DriverInstaller_v110.exe for installation the
driver.
2.The set up status.
3.Click “NEXT” to continue.
4.After complete the installation, press “Finished”.
6
7
B. Install the logger software for Windows XP
(win7/win8/VISTA)
Please follow the following steps to install the software
1. Put CD into CD driver, click “Setup.exe” to install
2.Select the destination folder you are go to save by click
Browse button
“
”
3.Once you finish selecting ,press Next
“
”.
4.Select “I accept the License Agreemen:(s)” and press NEXT
button.
5.Click NEXT button when you see the following.
6.Press FINISH button to complete the installation.
7.Restart the Windows XP system
C Run the software from Windows XP
.
1. Plug the USB cable into USB port, since you have already
Installed the USB driver, computer will automatically detect
logger ,
2. Find “Datalogger” in START PROGRAMS
3.Select logger software folder.
Reminder:
Software is installed completed,”Get Datalog’s Identifier”shows
next page(pic.1), then press SETTING button to enter Log setting
(Pic.1)
Note: If logger is not plugged in USB port, the warning message
“Reminder: Data Logger is not plugged in USB port!” shows next
page(pic.2) Make sure logger is plugged in and choose the
correct COM port.
MADE IN TAIWAN
AQM-101
IAQ Formaldehyde
(HCHO) Logger
M5443/1114
INSTRUCTION