3.  Recommended Position

4.  Installation

1

Preamplifier

2

Fitting
inlet

Hand 

tighten

only!

In-line Applications

Preamplifier

1

2

Customer 
supplied
pipe

3/4 in. NPT or
ISO 7/1-Rc 3/4

Submersible Applications

Sealant

PH-31542-TC Sensor Cap 
 (purchased separately)

A. O-ring lubricant
recommended
(factory applied)

A. O-ring lubricant
recommended
(factory applied)

Sealant

B. Align
marks

B. Align
marks

C.

C.

Fill with

3 to 4 in.

of sealant

3

•  Vertical (0

°

) position optimum 

•  4 fps or less for max. performance and sensor life 

0

°

 

Process 

pipe

+60

°

Customer supplied:
• 3/4 in. NPT threaded

pipe (A)

• J-box (B)
• Flex conduit (C)
• Quick release pipe 

clamp (D)

• Tank flange (E)

A.

B.

C.

C.

E.

B.

A.

D.

Closed

tanks

Open 

tanks

In-Line

Caution: If liquid level is not constant, always ensure 

liquid contact with electrode tip

Customer supplied:
•  3/4 in. NPT nipple (male) 
•  J-box
•  flex conduit

-60

°

3/4 in. NPT OR 
ISO 7/1-Rc 3/4
female threads

2 in. 

min.

Attach 3/4 in. watertight pipe to the top of the preamplifier.  Secure
the threaded connection to prevent any leakage.  For  additional
defense against possible accumulation of condensation at the back
seal area of the sensor, fill the lower 3-4 inches (75-100 mm) of
conduit or extension pipe with a flexible sealant such as silicone.

Summary of Contents for PH-2720-PA

Page 1: ...www omega com e mail info omega com User sGuide Shop online at PH 2720 PA Preamplifier...

Page 2: ...ance USA and Canada Sales Service 1 800 826 6342 1 800 TC OMEGA Customer Service 1 800 622 2378 1 800 622 BEST Engineering Service 1 800 872 9436 1 800 USA WHEN TELEX 996404 EASYLINK 62968934 CABLE OM...

Page 3: ...n Blue Green White Red Black mV Input Iso Gnd T T V V OMEGA PHCN 5700 OMEGA ORCN 90 ORP CONTROLLER ENTER MOD CAL RELAY OUTPUT ORP mV RLY 1 RLY 2 RLY 3 RLY 4 Red Black Green White Brown Blue 10 V 11 V...

Page 4: ...rocess pipe 60 Customer supplied 3 4 in NPT threaded pipe A J box B Flex conduit C Quick release pipe clamp D Tank flange E A B C C E B A D Closed tanks Open tanks In Line Caution If liquid level is n...

Page 5: ...stallation instructions see sensor manual 4 Wear safety goggles and faceshield during installation service 5 Do not alter product construction 6 Failure to follow safety instructions could result in s...

Page 6: ...sensor tip in pH 4 0 buffer during system maintenance intervals In line applications should be plumbed with a depression trap which ensures liquid is maintained around the sensor tip If sensor dehydr...

Page 7: ...shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA...

Page 8: ...ers Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pum...

Reviews: