background image

1.  Wiring

OMEGA PH-2720-PA pH/ORP Preamplifier Instructions

SAFETY INSTRUCTIONS  FOR IN-LINE ELECTRODE INSTALLATION
1.

Do not remove from pressurized lines.

2.

Do not exceed maximum temperature/pressure specifications.

3.

Wear safety goggles or face shield during installation/service.

4.

Do not alter product construction.

Failure to follow safety instructions may result in severe personal injury!

2.  Installation Fittings

Consult OMEGA for additional fitting information.

ENTER

0

ALARM 1

ALARM 2

2

4

6

7

8

10

12

14

pH

OMEGA

PHCN-5700

Brown
Blue
Black
Red
Green
White

(mV Input)

(Iso. Gnd)

(V-) 

(V+)

(T+)

(T-)

Silver (earth ground)

RLY 1
RLY 2

RLY 3

RLY 4

pH CONTROLLER

ENTER

MOD

pH

CAL

RELAY

OUTPUT

mV

TEMP

OMEGA

PHCN-90

Red 
Black
Green
Brown
White
Blue

(#10) V+  

(#11) V-  

(#12) T+ 

(#13) Ain

(#14) ISO

(#14) ISO

• Use X1 gain selection

Silver (earth ground)

Technical Notes:

• Use 6-conductor shielded cable for cable extensions to 120 m (400 ft)
• Shield must be maintained through cable splice

Silver (earth ground)

Brown
Blue
Green
White
Red
Black

(mV Input)

(Iso. Gnd) 

(T+)

(T-)

(V+)

(V-)

OMEGA

PHCN-5700

OMEGA

ORCN-90

ORP CONTROLLER

ENTER

MOD

CAL

RELAY

OUTPUT

ORP

mV

RLY 1
RLY 2
RLY 3
RLY 4

Red
Black
Green
White
Brown
Blue

(#10) V+ 

(#11) V-  

N/C

N/C

(#13) Ain

(#14) ISO

• Use X1 gain selection

Silver (earth ground)

Preamplifier:
PH-2720-PA

pH 
Electrodes:

PHE-3271
PHE-2716

ORP
Electrodes:

ORE-2715
ORE-2717

Preamplifier:
PH-2720-PA

pH 
Electrodes:

PHE-3271
PHE-2716

ORP
Electrodes:

ORE-2715
ORE-2717

Preamplifier:
PH-2720-PA

pH 
Electrodes:

PHE-3271
PHE-2716

Preamplifier:
PH-2720-PA

ORP
Electrodes:

ORE-2715
ORE-2717

• 0.5 to 4 in. 
• PVC or CPVC
• Mounts via glue-on fittings

Plastic tees

PVC glue-on
saddles
(O-ring not
required)

• 2 to 4 in., use 1-7/16 in. hole in pipe
• Align wedge arrows with saddle 
   arrows during assembly.

Type

Description

Iron 
strap-on
saddles

• 2 to 4 in., use 1-7/16 in. hole in pipe

Carbon steel 
weld-on
weldolets

• 2 to 4 in, use 1-7/16 in. hole in pipe
• Remove insert before welding
• Installed by certified welder only 
  

• 0.5 to 2 in. 
• Mounts on threaded pipe ends

Carbon steel 
threaded tees

Universal
pipe adapter
for large pipes

• For existing pipe fittings 2 in. and up
• External 1-1/4 inch NPT male threads

Type

Description

Summary of Contents for PH-2720-PA

Page 1: ...www omega com e mail info omega com User sGuide Shop online at PH 2720 PA Preamplifier...

Page 2: ...ance USA and Canada Sales Service 1 800 826 6342 1 800 TC OMEGA Customer Service 1 800 622 2378 1 800 622 BEST Engineering Service 1 800 872 9436 1 800 USA WHEN TELEX 996404 EASYLINK 62968934 CABLE OM...

Page 3: ...n Blue Green White Red Black mV Input Iso Gnd T T V V OMEGA PHCN 5700 OMEGA ORCN 90 ORP CONTROLLER ENTER MOD CAL RELAY OUTPUT ORP mV RLY 1 RLY 2 RLY 3 RLY 4 Red Black Green White Brown Blue 10 V 11 V...

Page 4: ...rocess pipe 60 Customer supplied 3 4 in NPT threaded pipe A J box B Flex conduit C Quick release pipe clamp D Tank flange E A B C C E B A D Closed tanks Open tanks In Line Caution If liquid level is n...

Page 5: ...stallation instructions see sensor manual 4 Wear safety goggles and faceshield during installation service 5 Do not alter product construction 6 Failure to follow safety instructions could result in s...

Page 6: ...sensor tip in pH 4 0 buffer during system maintenance intervals In line applications should be plumbed with a depression trap which ensures liquid is maintained around the sensor tip If sensor dehydr...

Page 7: ...shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA...

Page 8: ...ers Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pum...

Reviews: