OM-CP-THERMOVAULTX Series
Device Maintenance
Battery Replacement
Materials: 3/32 inch HEX Driver (Allen Key) and a Replacement Battery (OM-CP-BAT103) or any
9 V battery
5. Remove the cover from the device by unscrewing the two screws.
6. Remove the battery from its compartment and unsnap it from the connector.
7. Snap the new battery into the terminals and verify it is secure.
8. Replace the cover taking care not to pinch the wires. Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
Recalibration is recommended annually for all Omega data loggers. The Properties window in the OM-CP
Data Logger software displays the date of the last calibration and the date that the device is next due for
calibration. The OM-CP Data Logger Software can also be configured to send an on screen notification prior
to the calibration due date for each device. By default this is set to seven days prior to calibration due date
and can be changed by the user by going to the file tab in the OM-CP Data Logger software and clicking on
Options
. Select device and check “
Display popup notification when a device nears its next calibration
date
”. The user can then select the number of days before calibration due date to notify.