3

Product Overview

The OM-CP-THERMOVAULTX Series is an oven temperature recorder that consists of a OM-CP-QUADTCTEMP 
or OM-CP-OCTTCTEMP inside a thermal enclosure. The thermal enclosure acts as a buffer and allows the 
device to be placed in higher operating environments for certain durations of time. The chart outlines the 
maximum temperature and time duration to which the device can be exposed.

Installation Guide

Installing the Interface Cable

Insert the USB-A to micro USB cable (included) into a USB port. The drivers will install automatically.

Installing the Software

Insert the OM-CP Data Logger Software Flash Drive into an open USB port on a Windows PC. If the autorun 
does not appear, locate the drive on the computer and double click on 

Autorun.exe

. Follow the instructions 

provided in the Installation Wizard.

 

Device Operation

Manual Start

Click the 

Custom Start

 button on the Device panel, or 

right-click on the device and hover on the start selection, 
then chose custom start. Apply the options desired and 
select 

Start

. Once armed through the software, to activate 

the Manual Start, hold the recessed push button, adjacent 
to the interface cable plug, down for 10 seconds. To see 
the change in the status, hit 

Refresh Devices

 within the 

software.

Thermocouple Type

To change the thermocouple type:
1.  In the 

Connected Devices

 panel, click the device desired.

2.  On the 

Device

 tab, in the Information Group, click 

Properties

. Or, right-click the device and select 

Properties

 in the context menu.

3.  On the 

General

 tab, change the 

Thermocouple type

 in the drop down menu.

4.  Apply these changes, there will be a prompt to reset the device, select 

Yes

.

Please note that the same thermocouple type must be used on all of the channels.

OM-CP-THERMOVAULTX Series 

Manual Start Button

Summary of Contents for OM-CP-THERMOVAULTX Series

Page 1: ...Xxxxxx Xxxxx Xxxxxx Xxxxxxxx M0000 0016 INSTRUCTION SHEET Shop online at omega comSM e mail info omega com For latest product manuals www omegamanual info TM OM CP THERMOVAULTX SERIES 4 8 Channel Thermocouple Data Logger ...

Page 2: ...es the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start stores the dataset as it records while connected to the logger The status of the device will change to Running Waiting to Start or Waiting to Manual Start depending upon your start method 5 Disconnect the data logger from the interface cable and place it in the environment to measure 6...

Page 3: ...s provided in the Installation Wizard Device Operation Manual Start Click the Custom Start button on the Device panel or right click on the device and hover on the start selection then chose custom start Apply the options desired and select Start Once armed through the software to activate the Manual Start hold the recessed push button adjacent to the interface cable plug down for 10 seconds To se...

Page 4: ...er is stopped with the logger highlighted click Download You will be prompted to name your report 4 Downloading will offload and save all the recorded data to the PC Set Password To password protect the device so that others cannot start stop or reset the device 1 In the Connected Devices panel click the device desired 2 On the Device tab in the Information Group click Properties Or right click th...

Page 5: ...s Recalibration Recalibration is recommended annually for all Omega data loggers The Properties window in the OM CP Data Logger software displays the date of the last calibration and the date that the device is next due for calibration The OM CP Data Logger Software can also be configured to send an on screen notification prior to the calibration due date for each device By default this is set to ...

Page 6: ...300 C 572 F 30 350 C 662 F 25 GENERAL Memory All channels enabled 4 channel 524 032 readings per channel 8 channel 262 016 readings per channel Start Modes Software programmable immediate start or delay start up to 6 months in advance LEDs 1 per channel and 2 status LEDs Reading Rate 4 Hz up to 1 reading every 24 hours Calibration Digital calibration through software Calibration Date Automatically...

Page 7: ... a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installation or activity medical application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnif...