- 4 - 

2

 

Installation 

2.1  Installation of the binocular viewing head 

1)  Loosen the head lock thumb screw (

Fig. 1

) on the 

top of the microscope body. 

Note:

   

The thumb screw may not be pre-installed. Please 
find it in the same package of the microscope. 

2)  Remove the cap on the circular dovetail of the 

binocular viewing head. 

3)  Insert the binocular viewing head into the top of the microscope body; ensure that 

the dovetail is completely seated into the socket and tighten the thumb screw.   

Caution

:   

Do not release the viewing head from your hand grip until you are sure the viewing 
head is installed securely.   

2.2  Installation of the eyepieces   

1)  Remove the protective caps from the eyepiece tubes.   
2)  Insert the eyepieces into the eyepiece tubes.     

2.3  Installation of the glass filter   

1)  Swing out the color filter holder under the condenser.   
2)  Place the filter into the holder as shown in 

Fig. 2

, swing the holder in.       

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2.4  Installation of the camera 

(optional, may not included in your package)

   

1)  Take off the eyepiece from the eyepiece tube. 
2)  Insert the camera into the eyepiece tube, and then 

connect the camera to a computer via USB2.0 
cable. See 

Fig. 3

3)  The manual for the camera is on a CD (or mini CD). 

Refer to the manual to install the driver and 
software on to the computer. 

4)  The camera is optional and may have different 

Fig. 2 

Fig. 1 

Head Lock Thumb Screw 

Fig. 3 

Camera 

Color Filter 

Color Filter Holder 

Summary of Contents for M82ES Series

Page 1: ...User Manual Binocular Compound Microscope Model M82ES Series MicroscopeNet com...

Page 2: ...aution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 8 5 Optional Parts 9 6 Troubleshooting Guide 10 7 Dry Darkfield Condenser Installation Ope...

Page 3: ...t of direct sunlight high temperature or humidity and dusty environments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto th...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...5 Objective 18 Power Switch 6 Slide Holder 19 Brightness Intensity Dial 7 Mechanical Stage 20 Immersion Oil 8 Light Collector 21 Dust Cover 9 Microscope Base 22 Condenser Lock Thumb Screw 10 Eyepiece...

Page 6: ...lled securely 2 2 Installation of the eyepieces 1 Remove the protective caps from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 3 Installation of the glass filter 1 Swing out the...

Page 7: ...ve to touch a specimen slide when changing objectives 3 3 Adjusting interpupillary distance While observing with both eyes hold the left and right eyepiece tubes and slide the tubes in and out The int...

Page 8: ...iece to seat the observing position between the 40X and 100X objectives as shown in Fig 6 a 2 Place a drop of immersion oil on the slide cover as shown in Fig 6 b 3 Rotate the objective nosepiece to s...

Page 9: ...adjust the brightness although the brightness will be changed when it s adjusted When aperture is adjusted to smaller size the contrast will be increased and the depth of field will be increased as we...

Page 10: ...0 17 spring oil Condenser Abbe NA 1 25 w iris diaphragm Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double layer di...

Page 11: ...200X 400X 800X 1000X 2000X 5 Optional Parts The optional parts may be included in some models or sold separately 1 Darkfield Condensers Model Darkfield Condenser Numerical Aperture Objective Mounting...

Page 12: ...sed image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at t...

Page 13: ...e the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn the...

Reviews: