Specifications 

 

General 

Model 

M82ES series 

Total Magnification 

40X, 80X, 100X, 200X, 400X, 800X, 1000X, 2000X 

Viewing Head 

Binocular, inclined 45º, swiveling 360º 

Interpupillary Distance 

Sliding adjustment, 2-3/16

” ~ 2-15/16” (55mm ~ 75mm) 

Diopter Adjustment 

On both eyepiece tubes 

Eyepieces 

1 pair of wild field WF10X/18 

1 pair of wild field WF20X 

 

Nosepiece 

Revolving quadruple 

Objectives 

Achromatic DIN 4X/0.10 160/0.17 

                            10X/0.25 160/0.17 

                            40X/0.65 160/0.17(spring), 

                            100X/1.25 160/0.17(spring, oil) 

Condenser 

Abbe, NA=1.25, w/ iris diaphragm 
Rack and pinion adjustment 

Focus Mechanism 

Coaxial coarse and fine focusing knobs on both sides 

with rack stop 

Mechanical Stage 

Double layer, dimension: 4-1/2" x 4-15/16" (115mm x 125mm) 
X-Y translational range: 2-

13/16” x 1-3/16” (70 mm x 30 mm)

 

Illumination 

Transmitted, LED, variable intensity 

Camera (

optional

Refer to the cameras specifications 

Dry Darkfield Condensers (

optional

Refer to the Dry Darkfield Condensers Installation & 

Operation Instruction 

Power Supply 

AC/DC adapter, 7.5V/7.5W (UL approved) 

- Input: 100-240VAC, 50/60Hz, 0.5A Max 

- Output: DC7.5V, 1000mA   

Dimension 

9-1/16

” x 7-1/8” x 13” (23cm x 18cm x 33cm) 

Net weight 

7 lbs 2 oz (3.25 kg) 

 

 
 
 
 

Summary of Contents for M82ES Series

Page 1: ...User Manual Binocular Compound Microscope Model M82ES Series MicroscopeNet com...

Page 2: ...aution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 8 5 Optional Parts 9 6 Troubleshooting Guide 10 7 Dry Darkfield Condenser Installation Ope...

Page 3: ...t of direct sunlight high temperature or humidity and dusty environments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto th...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...5 Objective 18 Power Switch 6 Slide Holder 19 Brightness Intensity Dial 7 Mechanical Stage 20 Immersion Oil 8 Light Collector 21 Dust Cover 9 Microscope Base 22 Condenser Lock Thumb Screw 10 Eyepiece...

Page 6: ...lled securely 2 2 Installation of the eyepieces 1 Remove the protective caps from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 3 Installation of the glass filter 1 Swing out the...

Page 7: ...ve to touch a specimen slide when changing objectives 3 3 Adjusting interpupillary distance While observing with both eyes hold the left and right eyepiece tubes and slide the tubes in and out The int...

Page 8: ...iece to seat the observing position between the 40X and 100X objectives as shown in Fig 6 a 2 Place a drop of immersion oil on the slide cover as shown in Fig 6 b 3 Rotate the objective nosepiece to s...

Page 9: ...adjust the brightness although the brightness will be changed when it s adjusted When aperture is adjusted to smaller size the contrast will be increased and the depth of field will be increased as we...

Page 10: ...0 17 spring oil Condenser Abbe NA 1 25 w iris diaphragm Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double layer di...

Page 11: ...200X 400X 800X 1000X 2000X 5 Optional Parts The optional parts may be included in some models or sold separately 1 Darkfield Condensers Model Darkfield Condenser Numerical Aperture Objective Mounting...

Page 12: ...sed image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at t...

Page 13: ...e the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn the...

Reviews: