background image

 

 

www.microscopenet.com

 

 

20 

 

Note:   

There are 3 phase contrast objectives: 10X, 40X and 100X, and there are 3 
condenser ring plates: 10X, 40X and 100X. The corresponding objective and ring 
plate must work together, i.e. 

10X

 phase contrast objective must work with 

10X

 

condenser ring plate, and so on. 

2)  Centering the condenser ring plate 

a. Connect the power cord to the microscope and insert the plug into a power outlet. 
b. Turn the desired phase contrast objective into light path. 
c.  Choose the ring plate that corresponding to the phase contrast objective in light path 

and screw it on the condenser. 

d. Replace the bright field condenser with the phase contrast condenser with ring plate 

on. 

e. Remove one eyepiece from the microscope eyepiece 

tube and insert the centering telescope.   

f.  Turn the light of microscope on and observe from the 

telescope.   

g. Turn the top of the telescope (

Fig.24

) until the dark ring 

image is in focus. If the dark ring is hard to find, put a 
normal white print paper on the stage and under the 
objective then you will see the dark ring as shown in 

Fig.23

 

(a)

.     

h. If the bright ring is still obscure as in 

Fig.23

 

(b)

, raise or 

lower the condenser by adjusting the condenser 
focusing knob (or the microscope focusing knob if 
necessary) till the bright ring is in focus and the dark 
ring is visible. 

i.  If the two ring images are not coincided as shown in 

Fig.23 (c)

, adjust the two centering screws on the 

condenser ring plate till the two rings are coincided with each 
other as shown in 

Fig.23 (d)

j.  Remove the centering telescope and replace it with the 

eyepiece. 

3)  Performing the phase contrast observation 

After you center the ring plate, you can perform the phase 
contrast observation the same way as a normal bright field 
microscope. 

Note:   

When change to another phase contrast objective and 
corresponding condenser ring plate, the focusing and 
centering of bright ring and dark ring should be repeated 
following the procedures from 

2)-b to 2)-h

Tips:   

  Make the illumination as bright as possible. 

  The thinner the specimen, the better the image.

 

 

 

(a) 

(b) 

(d) 

 

(c) 

Fig.23 

Fig.24 

Summary of Contents for M8211 Series

Page 1: ...User Manual Trinocular Compound LED Microscope Model M8311 Series MicroscopeNet com...

Page 2: ...1 Components Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 9 5 Optional Parts 10 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Page 3: ...he container should be retained should the microscope ever requires reshipment 3 Keep the instrument out of direct sunlight high temperature or humidity and dusty environments Ensure that the microsco...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...al 3 Diopter Ring 11 Nosepiece 19 Condenser Lock Thumb Screw 4 Eyepiece Tube 12 Objective 20 Condenser Focus Knob 5 Viewing Head 13 Microscope Body 21 Condenser 6 Slide Holder 14 Coarse Focus Knob 22...

Page 6: ...s from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn t...

Page 7: ...on the photo tube 2 Insert the camera into the photo tube and then connect the camera to a computer via USB2 0 cable See Fig 4 3 The manual for the camera is on a CD or mini CD Refer to the manual to...

Page 8: ...imen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus Then use the fine focus knob to obtain a sharp image You may now switch to another magnification objective 3...

Page 9: ...objectives When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not le...

Page 10: ...procedures in 3 3 2 Install the camera by following the procedures in 2 7 3 Open image observing software to examine more details see camera s manual 4 If the image is not clear loosen the 2 set scre...

Page 11: ...d filter holder Rack and pinion focus adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with stop Stage Double layer mechanical stage Dimension 5 1 2 x 4 1 4 140mm x 110m...

Page 12: ...Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above A3520U 1600 x 1200 2 0MP A3530U 2048 x 1536 3 0MP A3550U 2592 x 1944 5 0MP A3590U 3488x2616 9 0MP...

Page 13: ...eter A191 Dry 0 7 0 9 37mm A191BD Oil 1 36 1 25 37mm A191BOIL new design Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring with iris diaphragm 37mm 3 Carrying Case Model Dimension Net Weight A187V...

Page 14: ...Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil Ffive positions 10 for 10X phase contrast objective 20 for 20X phase contrast objective 40 for 40X...

Page 15: ...the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cove...

Page 16: ...nsufficient or too bright The intensity of lamp is too low Adjust the intensity dial to increase the brightness The condenser is not in the right position Lower or raise the condenser slightly to the...

Page 17: ...erve 6 Move the condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X...

Page 18: ...piece b Install the phase contrast objectives onto the nosepiece following the steps in 2 3 2 Mounting the phase contrast condenser annular ring disk a Rotate the nosepiece and set the 4X objective in...

Page 19: ...nular ring disk being put at B position 6 Performing the phase contrast observation After centering the condenser ring plate perform the phase contrast observation the same way as a normal bright fiel...

Page 20: ...rk ring should be coincided with each other as shown in Fig 19 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f If the bright ring is still obscure...

Page 21: ...bjective do not let the immersion oil to contact with the dry objective lenses especially the 40X If the immersion oil is on the dry objectives lens please use the lens cleaning paper to clean the obj...

Page 22: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 23 a h If the bright ring is still obscure as in Fig 23 b raise or l...

Reviews: