24
d
e
3
Move the
portion shown in the picture up and down so that the
d portion of the stand is 71 mm, and tighten the stand height lock
knob e .
If there is a difference in height between the desk top surface
where the microscope frame is installed and the desk top
surface where U-MDOSV is installed, adjust the length of the
d portion of the stand. For example, if the height difference
is 5 mm (if the desk top surface of the microscope frame is
higher than the desk top surface of U-MDOSV), d will be 76
mm.
3
TIP
Remove the cap from the mount a on BX3-MDOE and the cap from
the mount b on U-MDOSV.
Return the vertical direction of U-MDOSV where the stand is attached
to the original direction. Insert b of U-MDOSV into the mount a on
BX3-MDOE completely to the end.
1
2
6
Attaching U-MDOSV
While holding the base c of the stand firmly with your hand, rotate
the mount ring d to secure U-MDOSV to BX3-MDOE.
After the mount ring
d
is rotated completely, make sure
that the stand is installed on the desk securely. Insert a
sheet of paper between the base of the stand and the
desk. If the paper touches three rubber legs, the stand is
installed securely. If the paper does not touch even one
rubber leg, perform
3
again.
3
NOTE
a
b
2
d
c
7
Attaching the observation tube
See "Attaching the observation tube" in "5-2 Side by side
observation system for 2 persons and system for 5 persons" (page
19).