21
Look through the eyepieces at the main observer position
1
and
display the pointer using the pointer color switching lever a . Use the
pointer control lever b to move the pointer to the center of the field of
view. (For detailed operating procedures of the pointer color switching
lever and the pointer control lever, see "3-3 Pointer operation" (page 13).)
1
Look through the eyepieces at the assistant observer position
2
. If
there is a deviation in the displayed pointer position, adjust the stand
height with the stand height lock knob c so that the pointer comes
to the center of the field of view.
Be sure to adjust the stand height.
If the stand height is not adjusted, following phenomena may
appear.
· The periphery areas of the field of view of the assistant
observer are cut off.
· The deviation in the field of view occurs between the main
observer and the assistant observer.
· The field of view of the assistant observer is colored.
The farther away from the center of the system you apply
the force to the observation tube, the larger the force to be
applied to the entire system becomes. Do not apply the
excess force on the observation tube carelessly.
2
TIP
NOTE
6
Adjusting the stand height
c
6
Adjustment order
Adjust the stand height in the following order.
2
2
2
2
1
Main
observer
Assistant
observer
Assistant
observer
b
a