Always move the light path selector lever
a
until it hits
the stop position.
For observation through the binocular light path, push in the light
path selector lever a till the stop position.
In this position, the whole light (100%) goes to the binocular tube.
For the camera light path, pull out the light path selector lever a till
the stop position.
In this position, a mirror is inserted into the light path so the whole
light for the right eyepiece sleeve (100%) is sent to the camera light
path.
NOTE
1
2
7
Selecting the Light Path (SZ61TR)
a
All of the auxiliary objectives are provided with the ESD
protection performance.
·
The 0.3X and 0.4X auxiliary objectives cannot be used in
combination with the standard base because their working
distances are too long.
·
The 0.5X auxiliary objective (110AL0.5X-2) also has a long
working distance, but this can be combined with the
standard base provided that the auxiliary sleeve (SZ2-ET)
is used. In this case, however, the microscope becomes
unstable and needs care against toppling it down.
· Screw the auxiliary objective firmly into the auxiliary objective mount
thread at the bottom of the microscope body.
TIP
NOTE
6
Using the Auxiliary Objectives
Auxiliary sleeve
SZ2-ET
Microscope body
Standard
base