background image

20

8-2  Stage Adapter Type 1 SZH-STAD1

The SZH-STAD1 is an adapter providing the similar function to the BX stage adapter type 1 (SZX-STAD1). The applicable 
stage is the BH2-SH horizontal knob stage.

1

Assembly

Use the tools shown in the table below.

Tool

Description

Allen wrench (for M4 screw)

Provided with SZH-STAD1

Hexagon socket screw (M4)

Provided with SZH-STAD1

Placing the notch  a  on the front side of the base, secure the screws 
(2 pcs.) using the Allen wrench to attach SZX-STAD1  b  to a base. 

Place the stage clamping knob  c  of BH2-SH (horizontal knob stage) 
on the front of the base, rotate the clamping knob  c  in clockwise 
direction to secure it, and attach BH2-SH.

TIP

1

2

b

2

c

a

b

1

STAD mount for ILLT SZX2-STADM is necessary for using the SZX2-ILLTQ/ILLTS LED transmitted light 
illumination base. For details, refer to the instruction manual for the SZX2-ILLTQ/ILLTS LED.

TIP

Downloaded from 

ManualsNet.com

 search engine

Summary of Contents for SZ61

Page 1: ...this manual thoroughly before operating the microscope Retain this instruction manual in an easily accessible place near the work desk for future reference For details of products included in the conf...

Page 2: ...Downloaded from ManualsNet com search engine...

Page 3: ...Zoom Magnification Stopper 8 4 Using the Eye Shades 8 5 Mounting the Eyepiece Micrometer Disk 9 6 Using the Auxiliary Objectives 10 7 Selecting the Light Path SZ61TR 10 8 Adjusting the Confocality of...

Page 4: ...Downloaded from ManualsNet com search engine...

Page 5: ...SAFETY PRECAUTIONS b a Hold the bottom of the stand with one hand while holding base column with the other hand to avoid tilting the microscope In case the specimen is damaged by erroneous operation...

Page 6: ...the instructions in the warning could result in bodily harm to the user and or damage to equipment including objects in the vicinity of the equipment NOTE Indicates that failure to follow the instruct...

Page 7: ...crew x 2 Accessory mount holes Eyepiece WHSZ10X WHSZ10X H WHSZ15X H WHSZ20X WHSZ20X H WHSZ30X H Body clamping knob Threaded for accessory mounting Microscope Body SZ61 SZ61 60 SZ61TR SZ51 SZ51 60 Stag...

Page 8: ...clamping screw Confocality adjustment clamping screw Page 11 x 2 Light path selector lever Page 10 Pulled out Camera path 100 Right eyepiece sleeve light path 0 Pushed in Right eyepiece sleeve light p...

Page 9: ...6 Adjust the interpupillary distance Page 6 1 2 Adjust the eyepiece diopter Page 7 This operation is not possible with the eyepieces without helicoids WHSZ10X 20X The operation is slightly difference...

Page 10: ...set the knob tension to a slightly higher level than the point where spontaneous drop occurs Hold the left and right focus adjustment knobs a with both hands fix the left knob and rotate the right kn...

Page 11: ...easy to observe specimen on the stage plate Set the zoom adjustment knob c for low magnification look through the eyepiece with eyepiece micrometer disk and rotate the focus adjustment knob d to bring...

Page 12: ...on each zoom adjustment knob The right zoom adjustment knob is used to limit the higher end of the magnification and the left knob is used to limit the lower end Using an Allen wrench loosen the clamp...

Page 13: ...ch it firmly Themicrometerdiskframemaybetightforcertainmicrometerdisks In this case turn the frame by holding the circumference with a light uniform force or by applying the frame against a rubber she...

Page 14: ...of the auxiliary objectives are provided with the ESD protection performance The 0 3X and 0 4X auxiliary objectives cannot be used in combination with the standard base because their working distance...

Page 15: ...monitor image After the adjustment tighten the clamping screws b and c Applicable cameras image pickup device size and field numbers The camera light path incorporates a 0 5X lens TIP 1 2 3 4 TIP a b...

Page 16: ...does not rotate smoothly The rotation tension of the knob is set too high Decrease the rotation tension to an optimum level 6 6 The microscope body drops spontaneously causing the focusing to be devia...

Page 17: ...eyepiece micrometer disk mountable Standard stand SZ2 ST Frame installation Mounting diameter 76 mm Focus adjustment Rack and pinion using ball guide Knob rotation tension adjustable Focusing stroke 1...

Page 18: ...5X 2 200 110AL2X 2 38 110AL0 62X 160 The working distances of the 110ALK0 3X and 0 4X can be varied according to the system The indicated magnifications 0 3X and 0 4X are the values when the working d...

Page 19: ...NOTE To reserve the ESD protection performance be sure to ground the equipment using a grounding lead wire having a 4 mm banana plug The eyepieces are provided with theft prevention screws which can...

Page 20: ...e microscope body can be installed in the opposite orientation to that shown in figure at the 180 rotated position TIP 3 Mounting the Stage Plate Mounting While applying the stage plate c to the stage...

Page 21: ...C mount clamping screw a and remove the C mount seat b Screw the C mount seat b into the camera c Place the C mount seat in the original position and tighten the clamping screw a Connect the cables an...

Page 22: ...sleeve when the SZ2 ST is used or the SZH P400 auxiliary column and the SZX R drop prevention ring also when the SZX2 ST or SZX2 illumination base is used 2 Applicable Bases and Restrictions 1 Introd...

Page 23: ...attach SZX STAD1 a to a base Attach U SRP or U SRG2 to SZX STAD1 a Attaching U SRP Place the positioning pin cylinder type c on the back side of the base and rotate the centering knob d in clockwise...

Page 24: ...1 Placing the notch a on the front side of the base secure the screws 2 pcs using the Allen wrench to attach SZX STAD1 b to a base Place the stage clamping knob c of BH2 SH horizontal knob stage on th...

Page 25: ...AD mount for ILLT SZX2 STADM is necessary for using the SZX2 ILLTQ ILLTS LED transmitted light illumination base For details refer to the instruction manual for the SZX2 ILLTQ ILLTS LED Stage plate mo...

Page 26: ...wipe the sliding surface on the cup stage and seat with a clean cloth Mount the stage plate Mount the specimen holder Clean the sliding surface periodically 1 2 3 4 TIP Before mounting remove dirt and...

Page 27: ...to the specimen holder to fix the container by pinching TIP Do not touch the sliding surface on the cup stage and seat with a hand If the sliding surface are contaminated wash with a neutral detergent...

Page 28: ...ributed by Manufactured by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan www olympus global com REP EC AX6611 10 Issued in March 2019 Downloaded from ManualsNet com search en...

Reviews: