11
Safety information
CAUTION
Risk of injury to the patient and/or the user
Using incompatible equipment may lead to injury of the patient
and/or the user as well as damage to the product.
• For information on compatible equipment, refer to the chapter
“Compatible equipment” on page 41.
CAUTION
Risk of injury to the user
There is a risk of injury to the user when using endoscopic
equipment in combination with laser equipment.
•
Use suitable protection measures (e.g., filters, protective
goggles).
NOTICE
Risk of damage to the product
The endoscope is a precise optical device. Careless handling
may damage the endoscope.
• Always handle the endoscope with care.
• Do not hold the endoscope by the distal end only.
• Do not bend the insertion tube.
• Always store and transport the endoscope in the supplied
protective tube.
• Remove the endoscope from the protective tube only before
direct use, manual cleaning or automated cleaning and
disinfection.