10
Safety information
•
Before each use, check the applied part classification of
the endoscopic equipment and especially the energized
endotherapy devices used during the procedure.
• Only use endoscopic equipment or endoscopic equipment in
combination with energized endotherapy devices that meet at
least the same requirements of applied part classification.
This is particularly important if a type CF applied part
endoscope is used. In this case a type CF applied part
energized endotherapy device should be used to minimize total
leakage current.
•
For the applied part classification, refer to the respective
instructions for use.
WARNING
Risk of injury to the patient
There is a risk of electric shock when using light sources with
noninsulated light-guide connectors.
• Only use light sources or light-guide cables that meet at least
the same requirements of applied part classification.
•
For the applied part classification, refer to the respective
instructions for use.
WARNING
Risk of injury to the patient
There is a risk of injury to the patient due to malfunction of the
equipment.
• Always have spare equipment available.
WARNING
Risk of injury to the patient and/or the user
There is a risk of injury to the patient and/or the user caused by
unauthorized repairs and product modification. Possible injuries
include mechanical injuries, electric shock, burns, and toxication.
• Do not attempt to repair or modify the product.
•
Repairs may only be performed by qualified servicing
personnel that has been authorized by Olympus.
• Contact an Olympus representative or an authorized service
center for repair information.