11. Place the probe in a defect-free area of the calibration standard near the 20 % OD
flaw, and then press the NULL foot switch.
12. Press the FULL NEXT key (
).
13. Scan the entire length of the tube.
The scan result is shown in Figure 5-217 on page 276.
Figure 5
‑
217 The scan of the entire length of the tube