Chapter 6 Troubleshooting
MAINTENANCE UNIT MU-1
Chapter 6 Troubleshooting
If the instrument is visibly damaged, does not function as expected or is found to
have other irregularities during the inspection described in Chapter 3,
“Preparation and Inspection”, do not use the instrument. Contact Olympus.
Problems that appear not to be malfunctions may be correctable by referring to
Section 6.1, “Troubleshooting Guide”. Should you fail to correct the problem
even after taking the described remedial action, stop using the instrument and
send it to Olympus for repair. Olympus does not repair accessory parts. If an
accessory part becomes damaged, contact Olympus to purchase a replacement.
6.1
Troubleshooting Guide
Irregularity
Description
Possible Cause
Solution
The pump does not
work.
The Power Cord is not
connected.
Connect the Power Cord to a Wall
Mains Outlet.
The Power Switch is not
turned ON.
Turn the Power Switch ON “
”.
The pump works but no
air comes out.
The endoscope is not
correctly connected to
the unit.
Connect the endoscope to the unit
correctly.
The unit is not correctly
connected to the
leakage tester, or the
leakage tester is not
correctly connected to
the endoscope.
Connect the unit to the leakage
tester correctly, and connect the
leakage tester to the endoscope
correctly.
Insufflation pressure is
weak.
Air leaks from a
connection section.
Confirm no air leaks from any
connection section.