8 Troubleshooting
Common problems
Problem
Cause and recommended action
Illumination does not light.
Illumination is not lit.
Press the LIGHT button (
).
The illumination is dim.
The distal end is dirty.
Wipe with a clean piece of gauze or a cotton swab (See "Cleaning the distal end"
The system cannot be
turned on.
The scope unit is not connected improperly.
Connect the scope unit properly.
The AC adapter or the battery is not connected.
Connect the AC adapter or the battery correctly (See "Preparing the power supply"
The Power button is not on.
Press the Power button (
).
A non-designated AC adapter or battery is connected.
Use the designated AC adapter or battery.
The system cannot be
turned off.
The main unit is damaged.
Remove the AC adapter or the battery from the main unit to turn off the power.
The image is not sharp.
The objective lens on the distal end is dirty.
Wipe with a clean piece of gauze or a cotton swab (See "Cleaning the distal end"
The external monitor is adjusted incorrectly.
Correctly adjust the external monitor.
The image brightness is
not optimum.
The automatic brightness control is set erroneously.
Perform the [BRT] button operation to adjust to the proper setting.
The objective lens on the distal end is dirty.
Wipe with a clean piece of gauze or a cotton swab (See "Cleaning the distal end"
The external monitor is adjusted incorrectly.
Correctly adjust the external monitor.
Color reproduction is poor.
The white balance is set improperly.
Re-adjust white balance.
The image is noisy.
The automatic brightness control is set erroneously.
Perform the [BRT] button operation to adjust to the proper setting.
The scope was incorrectly initialized.
Initialize the scope again.
Angulation cannot be
returned to its original atti-
tude.
The angulation lock is engaged
Return the angulation lock lever to release the angulation lock.
The operation load for
articulation is getting
heavier.
The angulation lock is engaged
Return the angulation lock lever to release the angulation lock.
The beep cannot be heard.
[BEEP ON/OFF] is set to [OFF].
Set [BEEP ON/OFF] in the [SETUP] menu to [ON].