Olympus FV1000                                                                         -4-                                                                  Short instructions 
 

5.

 

Changing objectives: 

Never change the objectives manually! 

To change objectives use the microscope control menu within the Olympus 
software or the panel box left of the microscope stand. 

6.

 

Changing the specimen: 

Before you change the specimen, drive down the objective nosepiece.  

Don’t forget to remove the old oil from the objective before you add new ones. 

7.

 

Lasers: 
 
Lasers for the lines 491, 532, and 559 nm have to be switched on manually. Read 
chapter 3 for details. 

8.

 

Data transfer: 

Never store any data on the microscope computer. 

There are three ways to transfer your data from the microscope computer: 

a.

 

Transfer through the MHH network. Save your data on hard disk P: of your 
department. 

b.

 

USB-sticks or movable hard disks connected via USB. The USB sockets are 
located at the back of the computer.´ 

c.

 

Writable CDs. This computer supports only CD-R, -RW. Other formats will 
not be recognized. Open the Nero Start Smart software and follow the 
instructions. 

 

Summary of Contents for Fluoview FV1000

Page 1: ...the microscope system For a detailed instruction in system operation refer to the Olympus user manual This instruction also contains important information on laser safety and general guidelines Conten...

Page 2: ...r eyesight may be permanently damaged A reflected laser beam is just as dangerous as a direct beam 3 Never deactivate the laser protection devices Read the Safety Guide in the Olympus User s Manual to...

Page 3: ...hese hours pick up a key in advance When you are leaving lock the microscope room and after 17 00 also look the main door of the facility Return the key to Wolfgang Posselt or to the mail box of the L...

Page 4: ...7 Lasers Lasers for the lines 491 532 and 559 nm have to be switched on manually Read chapter 3 for details 8 Data transfer Never store any data on the microscope computer There are three ways to tra...

Page 5: ...nm are controlled entirely through the Olympus software These lasers are switched on by selecting an acquisition protocol from the dye list in the Image Acquisition window or by directly clicking on...

Page 6: ...able to avoid condensation which will damage the microscope equipment 5 If you are using the CO2 open the gas bottle at the wall behind the microscope table Turn the main valve counter clockwise not m...

Page 7: ...microscope 1 Exit the control software 2 Shut down Windows 3 Turn the key switch of the 491 532 nm laser to off no 4 4 Turn the key switch of the 559 nm laser to off no 5 5 Switch off all buttons of...

Reviews: