Chapter 5 Reprocessing
Rotatable Clip Fixing Device
52
1.
Immerse the entire instrument in the detergent
solution for the time specified in manufacturer’s
instructions. If no time is specified, immerse for
between 5 minutes and 3 hours.
2.
Remove the instrument from the detergent solution.
Ultrasonic Cleaning
1.
Immerse the entire instrument in the ultrasonic
cleaner containing detergent solution.
2.
Clean ultrasonically for 30 minutes. For details on
operation of the ultrasonic cleaner, refer to the
instruction manual of the ultrasonic cleaner.
3.
Remove the instrument from the detergent solution.
Rinsing
•
After ultrasonic cleaning, rinse the
instrument thoroughly to remove residual
detergent. Residual detergent solution
could cause tissue irritation in the next
patient.
• Do not forcefully squeeze, wipe or scrub
the instrument. This could cause damage
to the instrument or result in reduced
performance.
1.
Rinse the instrument under clean running tap water.
2.
Confirm that no debris is left on the surfaces of the
instrument.
3.
Wipe the exterior of the instrument with a clean, dry
lint-free cloth.