Chapter 3 Preparation, Inspection and Operation
Rotatable Clip Fixing Device
18
3.1
Preparation
Equipment and Personal Protective Equipment
Prepare all equipment and personal protective
equipment which will be used with the instrument and
clip in accordance with their respective instruction
manuals. Appropriate personal protective equipment
may include: eye wear, a face mask, moisture-resistant
clothing and chemical-resistant gloves.
Spare Instrument and Clips
Always have a spare instrument and clips available.
Reprocessing Equipment
Prepare the reprocessing equipment as described in
Section 5.2, “Required Reprocessing Equipment” for
immediate reprocessing after use.
Equipment to be Used in an Emergency
Always have pliers and/or wire cutters ready to cut the
coil sheath, tube sheath and operation wire in case the
clip cannot be detached from the instrument.