Chapter 3 Preparation, Inspection and Operation
Rotatable Clip Fixing Device
30
•
When inserting the instrument into the
endoscope, hold it close to the biopsy
valve and keep it as straight as possible
relative to the biopsy valve. Otherwise,
the insertion portion could be damaged.
• Insert the instrument slowly. Abrupt
insertion could damage the endoscope
and/or instrument. If the insertion portion
of the instrument is damaged, the
rotation function will be impaired.
• As the distal end of the coil sheath is
thicker than other parts, resistance may
be felt when it passes through the
section near the biopsy valve. In this
case, do not advance it forcibly but gently
advance it upright with respect to the
biopsy valve. Otherwise, deformation of
the coil sheath may result.
1.
Confirm that the entire clip is retracted into the coil
sheath.
2.
Carefully insert the instrument into the biopsy valve
(see Figure 3.8).
Figure 3.8
Keep as straight as
possible.
Hold the
insertion portion
close to the
biopsy valve.
Biopsy valve