3. Adjustment procedure (CKX41)
Preparation for optical adjustment
Adjusting parfocality
Adjusting revolving axis
Adjusting left / right optical axis
Adjusting absolute optical axis
Checking exit pupil center
Checking stage tilt alignment
Adjusting moving force (tension)
See D-9.
Jigs : See D-10.
Procedure: See E-3 to E-6
in CH3-BI45 repair manual.
Jigs : See D-10.
Procedure: See E-8 to E-9
in CH3-BI45 repair manual.
Jigs : See D-9 and D-10.
Procedure: See E-10 to E-11
in CH3-BI45 repair manual.
Jigs : See D-9 and D-10.
Procedure: See D-14 (6-2) in CK30/40
repair manual.
See D-11 to D-12.
See D-17 in CK30/40 repair manual.
See C-6 to C-7.
*In adjustment of revoving axis and left/right optical axis , it is nesessary to use CK40-MVR (and
supplied slide-glass holder ) when moving a specimen. If not available, move it by hand.
D-8
CKX31/CKX41
D. REPAIR PROCEDURE
See D-10.(adjustment in binocular tube)
(center of standard objective)