Part
Item
Standard
Method
Optical axis,
CH3-BI45 repair manual
Pafocality
Binocular
tube
<CKX41>
(Refer to E-2 - E-12 of
repair procedure)
CH3-BI45 repair manual
(Refer to E-13 - E-16 of
repair procedure)
Set the following jigs to the microscope frame
* 1) Since the binocular section of CKX41 is the same as that of CX31, basic procedure of optical axis adjustment is
performed in the same manner as CX31, which is referred to CH3-BI45 repair manual regarding the adjustment
2) Nesessary jigs are partially different because UIS optical system is adopted in CKX. (Refer to the above description.)
method.
exit pupil center
and check the optical axis and exit pupil center.
(1) Revolving axis
1) Standard eyepiece: KN0048 (with adapter-1)
2) Objective: PlanC10x
3) Test plate: concentric circles (KN0003) or
cross micrometer (OB-M1/100SQ)
referring the CH3-BI45 repair manual.
(2) Left/right optical axis
Same as the above
(3) Absolute optical axis
1) Standard eyepiece: KN0048 (with adapter-1)
2) Standard objective: KN0041
(4) Exit pupil
1) Centering telescope (KN0029)
2) Objective: PlanC10x
Set the following jigs to the microscope frame
and check the parfocality referring the CH3-BI45
repair manual.
1) Standard eyepiece: KN0048 (with adapter-1)
2) Standard objective: KN0041
3) Focusing telescope: FT-36 or U-FT
B-3
CKX31/CKX41
B. INSPECTION STANDARD