background image

21

BX61

3-6  Immersion Objectives

1

    Using Immersion Objectives

(Fig. 36)

1. Focus on the specimen using all objectives, starting from the lowest-

power objective to higher-power objectives.

2. Press the stage escape button to lower the stage, then place a drop of

provided immersion oil onto the specimen at the area to be observed.

3. Turn the revolving nosepiece to engage the immersion objective.
4. Press the stage escape button to raise the stage, then turn the focus

adjustment knob to focus on the specimen.

# Since air bubbles in the oil will affect the image quality, make sure

that the oil is free of bubbles.

a. To check for bubbles, remove the eyepiece and fully open the field and

aperture iris diaphragms, then look at the exit pupil of the objective inside
the observation tube. (The pupil should appear round and bright.)

b. To remove bubbles, turn the revolving nosepiece toward the shorter ob-

jectives and  repeatedly defocus and refocus the oil immersion objective.

}If the condenser engraving shows a numerical aperture (NA) of 1.0 or

higher, the number applies only when oil is applied between the slide
glass and the top surface of the condenser. When oil is not present, the
NA is about 0.9.

5. After use, remove oil from the objective front lens by wiping with gauze

slightly moistened with absolute alcohol.

!

Caution in use of immersion oil

If immersion oil enters your eyes or contacts with your skin, immediately
take the following treatment.

Eyes: Rinse with fresh water (for 15 minutes or more).
Skin: Rise with water and soap.

If the appearance of the eyes or skin is altered or pain persists, immedi-
ately see your doctor.

3-7  Objectives with Correction Collar

}If the cover glass thickness is not 0.17 mm, the objectives cannot mani-

fest their performances. If a correction collar equipped objective is used
in this case, the difference in thickness can be compensated for by ad-
justing the collar.

Adjustment Procedure

 · If the cover glass thickness is known, set the correction collar @ to that

value. (Fig. 37)

 · If the cover glass thickness is unknown, adjust the correction collar @

and focus adjustment knob alternately until the positioning with the highest
resolution is obtained.

# Be careful not to touch the correction collar @ when turning the

revolving nosepiece.

Fig. 36

Fig. 37

@

# Be sure to use the provided Olympus Immersion oil.

Summary of Contents for BX61

Page 1: ...copes Models BX61 and BX62 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before o...

Page 2: ......

Page 3: ...5 Selecting the Light Path of the Trinocular Tube 6 Adjusting the Tilt 1 3 Correct assembly and adjustments are indispensable to allow the microscope manifest its full performance If you are going to...

Page 4: ...ccordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the...

Page 5: ...take the infection prevention measures The product becomes unstable if its height is increased by an accessory mounted on it In this case take anti toppling measures to prevent the specimen from being...

Page 6: ...e touched with bare hands Before use carefully read the instruction manual Improper use could result in personal injury to the user and or damage to the equipment Indicates that the main switch is ON...

Page 7: ...solute alcohol is highly flammable it must be handled carefully Be sure to keep it away from open flames or potential sources of electrical sparks for example electrical equipment that is being switch...

Page 8: ...re iris diaphragm ring Page 20 Filter mount Page 9 Field iris diaphragm ring Page 19 Brightness adjustment buttons Page 8 Light path selector knob Page 17 Condenser height adjustment screw Page 19 All...

Page 9: ...cape button Page 12 F C button Page 11 Light Preset switch Page 8 Lamp ON OFF switch Stage UP button Page 12 Stage DOWN button Page 12 Reflected lamp housing Motorized fluorescence illuminator BX RFAA...

Page 10: ...n the light path Revolving nosepiece button Bring the specimen in focus Stage UP DOWN button focus adjustment knob Adjust the interpupillary distance Adjust the diopter Adjust the light axis Binocular...

Page 11: ...icroscope Set the Z board or AF board For the setup and installation of the Z board refer to page 30 For the information regarding the AF board refer to the instruction manual of the AF unit Use the A...

Page 12: ...igs 4 5 1 Press the Light Preset switch The PHOTO indicator lights up 2 Press one of the light intensity adjustment buttons to set the de sired brightness 3 Press the Light Preset switch again The ind...

Page 13: ...the U FC filter cassette Even when a filter cassette is used a filter with thickness of no more than 3 mm can still be placed on the filter mount Using the Filter Cassette Figs 8 11 Loading Filters in...

Page 14: ...ten the filter cassette When the filter cassette is installed the condenser may hit the filter cassette when the stage is lowered Therefore exercise caution when lowering the stage with the filter cas...

Page 15: ...the opposite side to the stage feed knobs 1 Loosen the clamping screw and remove the focus adjustment knob 2 Remove the seal from the focus adjustment knob screw hole on the other side and attach the...

Page 16: ...used The upper limit of stage movement has been set so that the stage stops within the coverage range of the damper spring of objective when a slide specimen thickness 1 2 mm is used When observing a...

Page 17: ...stage by sliding each slide from the front 3 After placing the slides as far as they will go gently release the clamping lever Microscopy with Single Slide Holder Fig 18 The specimen slide can easily...

Page 18: ...Rubber Stage Knob Caps Optional When the X Y axis knobs are fitted with the rubber caps the knobs can be adjusted without slipping and fine adjustment is possible by holding the knobs with a very lig...

Page 19: ...lower the stage to the lower limit then remove the stage from the microscope 2 Using an Allen screwdriver loosen the substage bracket clamping screw and remove the substage 3 Press the stage UP butto...

Page 20: ...defined double crossline can be seen in the field of view Figs 25 26 2 Looking through the right eyepiece rotate the coarse and fine focus ad justment knob to bring the specimen and double crossline...

Page 21: ...ounterclockwise to remove it from the eyepiece and place a micrometer disk Screw the micrometer mounting frame back into the eyepiece sleeve Fig 28 Fig 29 5 Fig 29 Trinocular tube Pushed in Middle pos...

Page 22: ...r no intermediate attachment can be combined when the BX RFAA or BX RLAA reflected light illuminator is used If you need photography using the U TBI3 also use the U TRU trinocular intermediate tube Wi...

Page 23: ...o condenser centering screws and turn them to move the iris diaphragm image to the center of the field of view 6 Gradually open the field iris diaphragm The condenser is properly cen tered if the iris...

Page 24: ...ure iris diaphragm ring Fig 35 2 Compatibility of Objectives and Condensers Fig 34 Fig 35 N A scale Objective Magnification Condenser Achromat aplanat U AAC Swing out U SC3 Motorized universal U UCD8A...

Page 25: ...nly when oil is applied between the slide glass and the top surface of the condenser When oil is not present the NA is about 0 9 5 After use remove oil from the objective front lens by wiping with gau...

Page 26: ...ht path selector knob is in an posi tioned halfway Set the knob according to the obser vation method 17 Revolving nosepiece is not correctly en gaged Make sure that the revolving nose piece clicks pro...

Page 27: ...olving nosepiece is not corrected mounted Push slide dovetail all the way until it is stopped Objective is not correctly engaged in light path Make sure that revolving nosepiece clicks into place corr...

Page 28: ...h sides are the same type Your view is not accustomed to micro scope observation Upon looking into eyepieces try look ing at overall field before concentrat ing on specimen range You may also find it...

Page 29: ...1 mm fine 1 mm coarse Full stroke range 25 mm to 14 mm when the U FWT filter wheel is used 4 Revolving nosepiece 5 Observation tube Type U BI30 2 U TBI3 U TR30 2 U SWTR 3 Widefield bin ocular tube Wi...

Page 30: ...6 5 super widefield 2X 100X for FN 26 5 super widefield 8 Operating environment Indoor use Altitude Max 2000 meters Ambient temperature 5 to 40 C 41 to 104 F Maximum relative humidity 80 for temperatu...

Page 31: ...us m Field of view PlanN P Plan Achromat for polarized light FN22 10X 0 25 6 0 1 30 100X 28 0 2 2 20X 0 4 3 0 0 17 0 84 200X 9 3 1 1 40X 0 65 0 45 0 17 0 52 400X 2 0 0 55 100XO 1 25 0 13 0 27 1000X 0...

Page 32: ...35 1000X 1 0 0 22 Correction collar PlanFl Plan Semi Apochromat FN26 5 20X3 340 0 75 0 55 0 17 0 45 200X 4 3 1 1 20XW3 340 0 70 0 40 0 17 0 48 200X 4 08 1 1 40X3 340 0 90 0 20 0 11 0 23 0 37 400X 2 0...

Page 33: ...rformance we recommend that you have your Olympus representative assemble or remove this module Eyepiece WHN10X FN 22 35WHN10X FN 22 SWH10X H FN 26 5 35SWH10X FN 26 5 For the electrical connections re...

Page 34: ...on Mounting the Z Board Fig 39 Be sure to set the main switch on the BX UCB control box to OFF before proceeding to mounting the Z board 1 Loosen the 6 clamping knobs on the 2 option slot covers on th...

Page 35: ...ove Tighten the clamping screw while pressing downward on the cover Fig 40 Caution for Bulb Replacement During Use or Right After Use The bulb lamp socket and areas near these will be extremely hot du...

Page 36: ...sen the condenser clamping screw using the Allen screw driver When attaching the U SC3 swing out condenser align the positioning pin at the back of the condenser with the groove in the condenser sleev...

Page 37: ...ing Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The...

Page 38: ...ter sterreichischer Verband f r Elektrotechnik VE Danmarks Elektriske Materialkontroll DEMKO National Standards Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electrote...

Page 39: ......

Page 40: ...ny 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A One Corporate Drive Orangeburg NY 10962 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilb...

Reviews: