background image

Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-6RE Modular Revolving Nosepiece               

   

 

Page 9 of 18 

 

 

Revision 3 

 

Figure 13 

 Remove the pivot-adjustment screw 

Remove the Center-Pivot Bearing Ball 

Use a tweezers to remove the ¼” bearing ball from the 

center bore in the stationary base (se

Figure 14

).

 

 

Figure 14 

 Remove the center-pivot bearing ball 

Remove the Mechanical Detent Stop 

The next step is to remove the mechanical detent stop.  
Use a suitable JIS screwdriver to remove the two M2X4 
pan-head  screws  securing  this  stop  to  the  stationary 
base  of  the  turret  assembly  (see 

Figure  15

).    These 

screws  are  staked  in  place  with  adhesive  and  can  be 
surprisingly  stubborn  to remove, so be  sure  to use  the 
proper  JIS  screwdriver  to  prevent  damaging  the  screw 
heads.  It might also be helpful to heat the screws with 
a heat gun before loosening them, but do not melt the 
black knurled-rubber grip ring in the process. 

 

Figure 15 

 Remove the screws securing the detent stop 

Remove  the  loose  mechanical  detent  stop  from  the 
stationary base (see 

Figure 16

). 

 

Figure 16 

 Remove the mechanical detent stop 

Remove the Knurled-Rubber Grip Ring 

Carefully  remove  the  black  knurled-rubber  grip  ring 
from  the  outer  perimeter  of  the  revolving  turret  (see 

Figure 17

), being careful not to stretch or damage it in 

the process.  This ring is held onto the turret with a few 
spots of adhesive, so proceed carefully here.

 

 

Figure 17 

 Remove the knurled-rubber grip ring 

Summary of Contents for BH2-6RE

Page 1: ...tribute this PDF document in its entirety for personal non commercial purposes only provided that the contents are not modified in any way including the copyright notices contained herein This documen...

Page 2: ...rimeter Bearing Balls 10 Remove the Stationary Base from the Turret 10 Clean Grease from the Various Components 11 Reinstall Stationary Base into Revolving Turret 11 Apply Ring of Grease for the Perim...

Page 3: ...ure 18 Components of the turret assembly 10 Figure 19 Loosen the threaded retaining ring 10 Figure 20 Remove the threaded retaining ring 10 Figure 21 Remove the perimeter bearing balls 10 Figure 22 Li...

Page 4: ...f 18 Revision 3 Figure 43 Reinstall the knurled rubber grip ring 15 Figure 44 Remove any silicone RTV squeeze out 15 Figure 45 BH2 6RE ready to be put back into service 15 Figure 46 Revolving turret w...

Page 5: ...on revolving modular nosepiece assembly used on the Olympus BHS BHSP BHT and BHTP microscope stands This procedure is also applicable the BH2 5RE five position modular revolving nosepiece assembly due...

Page 6: ...due to solvent absorption The list of solvents generally considered safe for nylon includes acetone diethyl ether heptane mineral spirits naphthalene and turpentine Before Starting the Overhaul Remov...

Page 7: ...stment screw see Figure 8 can be difficult to remove unless the proper tool is used Do not attempt to remove the slotted lock ring and pivot adjustment screw unless you have access to such a tool sinc...

Page 8: ...ds in the revolving turret Figure 10 Insert screwdriver into bores to lock turret Loosen and Remove the Slotted Lock Ring Hold the turret assembly by grasping the knurled rubber grip ring on the outer...

Page 9: ...lace with adhesive and can be surprisingly stubborn to remove so be sure to use the proper JIS screwdriver to prevent damaging the screw heads It might also be helpful to heat the screws with a heat g...

Page 10: ...ter punch or nailset tool into one of the six detent notches in the revolving turret instead of into one of the four tooling holes in the threaded retaining ring or the revolving turret will be irrepa...

Page 11: ...er of the revolving turret see Figure 24 Do not apply too much grease here to minimize the squeeze out that will occur when the threaded retaining ring is reinstalled to hold the stationary base into...

Page 12: ...that the center pivot bore has been greased reinstall the bearing ball into the freshly greased center pivot bore The grease will hold the center pivot bearing ball in the proper position during subs...

Page 13: ...oosen the slotted lock ring readjust the pivot adjustment screw and retighten the slotted lock ring as described above until the turret motion feels acceptable Clean Off Any Visible Grease Use a suita...

Page 14: ...the stationary base see Figure 40 Be sure to orient the dovetail slide such that the relief notch see inset of Figure 40 is facing downwards and towards the slotted lock ring on the pivot adjustment s...

Page 15: ...removing the protective cover cleaning and re greasing the detent notches and then reinstalling the protective cover per the procedures detailed in this document The dovetail slide does not need to be...

Page 16: ...n properly adjusted to remove the play in the center pivot mechanism How to Contact the Author Please feel free to direct any questions or comments regarding this document or BH 2 microscopes in gener...

Page 17: ...resistant layer of dull silver colored plating on the exposed surface When the hard plating in the area of the mechanical detents begins to wear away as a result of severe use without routine cleaning...

Page 18: ...t Item Description Manufacturer Manufacturer Model Vendor Vendor Price S1 E6000 adhesive black 2 oz 32329 Amazon 8 22 S2 Plastilube Brake Grease 75 cc Plastilube ATE70015 Amazon 10 88 Plastilube Brake...

Reviews: