Figure 5
‑
14 Measuring with the manual echo
‑
to
‑
echo detection mode
In severe corrosion situations where valid multiple echoes are not present, you must
use the standard mode to be able to measure thicknesses.
You can use dual element transducers in all three modes. You can use all
measurement, display, and datalogger functions with the echo-to-echo modes. The
internal datalogger recognizes and saves all necessary echo-to-echo information to
upload and download thickness, waveform, and setup data.
You do not have to switch between echo detection modes when measuring both
coated and uncoated areas because you can measure uncoated wall thicknesses using
an echo-to-echo mode.
To change the echo detection mode
1.
Press
[2nd F]
,
[ZOOM] (E
‑
TO
‑
E)
.
Dual element (DE)
manual echo-to-echo
(MEtoE) detection mode
E1 blank bar
Echo-to-echo detection
bar
[WAVE ADJ]
parameter
NOTE
TIP
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...