OLYMPIC BILI-METER
Instruction Manual
5
Taking Measurements
To use the Type B-22 sensor:
1
Connect the sensor to the readout.
2
Hold the sensor against the infant’s body as near to the umbilical as possible
and aim the sensor at the center of the phototherapy light (see Figure 7).
For reproducible measurements, always hold the sensor on the same place on
the infant’s body.
Changes in the distance or angle of the light to the patient will change the
irradiance the patient receives, requiring new measurements to be taken.
Figure 7
Measuring with the Type B-22 sensor
3
Press and hold
READ
. “On” briefly displays, then the irradiance measurement
displays. While pressing
READ
, adjust the aim of the sensor to obtain the
maximum reading.
4
Release
READ
. The irradiance measurement locks on the display for 30 seconds
then automatically shuts off to preserve battery life.
When charting irradiance, note the type of radiometer used (e.g., Olympic
Bili-Meter) and sensor (e.g., Type B-22).
If more than one phototherapy light is being used on the infant, take separate
measurements for each light and chart each reading.
Phototherapy light
Sensor aimed for maximum reading