Guidance and Manufacture’s Declaration 

- Electromagnetic Immunity 

The Versatile Table is intended for use in the electromagnetic environment specified below. The customer or the user should 
assure that it is used in such an environment. 

IMMUNITY Test 

IEC 60601 Test Level 

Compliance 

Level 

Electromagnetic Environment - Guidance 

 
Conducted RF 
IEC 61000-4-6 
 
 
 
Radiated RF 
IEC 61000-4-3 

 
3 V

rms

 

150 kHz to 80 MHz 
 
 
 
3 V/m 
80 MHz to 2,5 GHz 

 
3V

rms

 

 
 
 
 
3 V/m 

Portable and mobile RF communications equipment should 
not be used closer to any part of the Versatile Table, 
including cables, than the recommended separation 
distance calculated by the equation applicable to the 
transmitter frequency. 
Recommended separation distance 

 
 
80 MHz to 800 MHz 
 
800 MHz to 2,5 GHz 
 
 

Where 

P

 is the transmitter maximum output power rating in 

watts (W) according to the transmitter manufacturer and 

d

 

is the recommended separation distance in meters (m). 
Field strengths from fixed RF transmitters, as determined 
by an electromagnetic site survey 

a

, should be less than 

the compliance level in each frequency range. 

b

 

Interference may occur in the surroundings of equipment 
marked with the following symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects, and people. 
 

 
 
 
 
 
 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measure filed strength in the location in which the Versatile Table is used exceeds the applicable RF 
COMPLIANCE LEVEL above, the Versatile Table should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the Versatile Table. 
 
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 
 

Recommended Separation Distances Between Portable and Mobile RF 

Communications Equipment and the Versatile Table 

The Versatile Table is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The 
customer or the user can help prevent electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the Versatile Table as recommended below, according to the 
maximum output power of the communications equipment. 

Rated Maximum Output 

the Power of the Transmitter 

(W) 

Separation Distance According to Frequency of Transmitter (m) 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2,5 GHz 

0,01 

0,12 

0,12 

0,23 

0,1 

0,37 

0,37 

0,74 

1,2 

1,2 

2,3 

10 

3,7 

3,7 

7,4 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can 
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of 
the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects, and people. 

 

Summary of Contents for VERSATILE

Page 1: ......

Page 2: ......

Page 3: ...Operator and Others 8 5 5 Emergency Position 8 5 6 Paper Roll Holder Optional 9 5 7 Sliding Extender 9 5 8 Upholstered Leg Rest 9 5 9 Left and Right Bracket Locks 10 5 9 1 Accessories on Bracket Locks...

Page 4: ...15 10 Technical Features 15 10 1 Electromagnetic Compatibility EMC 16 11 Dimensional 18 12 Symbology 19 13 Important Notes 19 13 1 General Cares Compulsory Reading 20 13 2 Disposal 21 13 3 Transport a...

Page 5: ...ge Contents Check out the equipment package contents Standard Items 1 Versatile Table 1 Left Bracket Lock 1 Right Bracket Lock 2 Upholstered Leg Rest 1 Versatile Quick Operation Guide Optional Items 1...

Page 6: ...wing items of the Versatile Table are considered as applied parts to the patient Standard Items Upholstery Upholstered ler rest Seat extension Extender upholstery Optional Items Auxiliary pillow Uphol...

Page 7: ...lso it has a fuse holder for equipment 2 an auxiliary socket 3 and a fuse holder for this socket 4 To turn Off the equipment press the On Off switch 1 to the O position The headrest 1 was designed to...

Page 8: ...0 centimeters remaining out of the equipment movement area and its components To ensure safe and proper patient positioning the operator should advise him to keep his hands in sight during the entire...

Page 9: ...sliding extender 2 until its structure engages on the latches 3 To use the stainless tray 4 remove the extender upholstery 1 To use the upholstery 1 position it on the latches 3 To retract the sliding...

Page 10: ...sh it upwards causing the lever to pass through the metallic knob s axis 5 and then rotate it again 5 9 1 Accessories on Bracket Locks Accessories which can be installed on the Versatile Table by usin...

Page 11: ...optimizing the professional s work space This accessory can receive Colposcopes with pin fitting up to 22 5 mm diameter 7 8 It has arm 3 with free joint allowing better positioning of the Colposcope...

Page 12: ...red with non slip coating it assists in the patient s boarding and allows the professional to better approach the equipment since the step 1 can be repositioned inside the cabinet For the professional...

Page 13: ...tion systems that act on the mechanical and electrical parts of the set Pressurized suspension the high pressure nitrogen cylinder relieves the load of the gear motor lifting system of the equipment i...

Page 14: ...proper functioning of the equipment and its components read Chapter 5 Description and Operating of the Equipment of this manual Installation should only be made by an authorized technician Installatio...

Page 15: ...Power 118 127 V 200 VA Power 220 230 V 250 VA Protection fuses For 220 230 VC 1 5 A H 5 x 20 mm For 118 127 VC 3 A H 5 x 20 mm Power cable specification in compliance with requirements 6 1 and 6 2 of...

Page 16: ...nment IMMUNITY Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wo...

Page 17: ...environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measure filed strength in the location in which the Versatile Table is used exceeds the applicable...

Page 18: ...18 11 Dimensional...

Page 19: ...its publication Future technical improvements do not result in any right to update existing products The images presented in this manual are illustrative This equipment is designed to be free from in...

Page 20: ...s not suitable for use in the presence of a flammable anesthetic mixture with air O2 or Nitrous Oxide This equipment should only be operated by physicians and nurses for clinical outpatient and gyneco...

Page 21: ...s for transport and storage 20 to 70 This equipment is contraindicated for any use other than that for which it is intended or to be operated by unauthorized personnel 14 Troubleshooting For solving p...

Page 22: ...ck the left and right bracket locks OPTIONAL ITEMS Check Colposcope holder joints Check the double bracket locks Check the armrests Check the knees support bracket lock Check heels support bracket loc...

Page 23: ...Connection to wrong voltage power supply 5 This warranty will cease when a In the normal course of its period of warranty b Make changes in the product not authorized by Olsen c Adulterations in the p...

Page 24: ...nd of low maintenance cost These characteristics have been achieved thanks to our competent and dedicated team which make me very proud for many reasons as they are always giving the best of their cre...

Page 25: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 26: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 27: ......

Page 28: ...55 48 2106 6000 export3 olsen odo br Registration at Ministry of Health of Brazilian 10281300011 Techinician In Charge Eng Cleber da Costa CREA SC S1 116283 5 Cod 5400400 Rev 05 17 05 2019...

Page 29: ...VERSATILE...

Page 30: ......

Page 31: ...8 5 5 Posici n de Emergencia 8 5 6 Soporte para Rollo de Papel Desechable Opcional 9 5 7 Extensi n Deslizante 9 5 8 Pierneras Tapizadas 9 5 9 Abrazaderas Izquierda y Derecha 10 5 9 1 Accesorios sobre...

Page 32: ...cnicas 15 10 1 Compatibilidad Electromagn tica EMC 16 11 Dimensional 18 12 Simbolog a 19 13 Notas Importantes 19 13 1 Cuidados Generales Lectura Obligatoria 20 13 2 Descarte 21 13 3 Transporte y Almac...

Page 33: ...del Empaque A continuaci n observe el contenido del empaque Art culos Standard 1 Mesa Versatile 1 Abrazadera Izquierda 1 Abrazadera Derecha 2 Pierneras Tapizadas 1 Gu a R pida de Operaci n Mesa Versat...

Page 34: ...oto 15 Cajones laterales 16 Abrazadera Derecha Se consideran partes aplicables al paciente los siguientes art culos Art culos de Serie Tapizados Pierneras tapizadas Opcionales Almohada auxiliar Apoyab...

Page 35: ...ctrico tambi n est n situados el porta fusibles de protecci n del equipo 2 una toma auxiliar 3 y el porta fusibles de protecci n 4 de esta toma Para apagar el equipo presione el interruptor de On Off...

Page 36: ...a del rea de movimiento tanto del equipo como de sus componentes La posici n segura y adecuada del paciente sobre el equipo la debe orientar el operador para que el paciente mantenga se con sus manos...

Page 37: ...oclave de vapor Para acceder a estos elementos tire la extensi n deslizante 2 hasta que su estructura se encaje en las trancas Para utilizar la bandeja de acero inoxidable 4 retire el tapizado extenso...

Page 38: ...arriba haciendo que la varilla met lica 4 pase por dentro de su eje 5 para luego girarla nuevamente 5 9 1 Accesorios sobre Abrazaderas Accesorios fijos en la Mesa Versatile por medio de abrazaderas C...

Page 39: ...io para el trabajo del profesional Este accesorio puede recibir Colposcopios con pin de acople de hasta 22 5 mm de di metro 7 8 Presenta el brazo 3 con articulaci n libre lo que permite una mejor posi...

Page 40: ...e en 24 horas la reducci n bacteriana es superior a 99 9 Carenados producidos en ABS de alta resistencia con cobertura en acr lico no necesitan de pintura lo que facilita el pulido superficial para re...

Page 41: ...os para conectar el equipo a la red Si se han comprado accesorios opcionales estos tambi n deben instalarse en este momento El t cnico tambi n inspeccionar el equipo asegur ndose de que est de acuerdo...

Page 42: ...sen no aconseja el uso de cualquier tipo de producto qu mico para la limpieza de estas partes sin embargo para desinfecci n es importante verificar si el producto presenta compatibilidad y caracter st...

Page 43: ...cho 10 A 250 V NBR 14136 Tipo de protecci n contra choque el ctrico equipo Clase I de acuerdo con IEC 60601 1 y IEC 60601 1 2 Grado de protecci n Tipo B Modo de operaci n no continuo Tiempo On 30 s Ti...

Page 44: ...ticas IEC 61000 4 2 6 kV por contacto 8 kV por aire 6 kV por contacto 8 kV por aire Los pisos deber an ser de madera u hormig n o estar provistos de baldosas de cer mica Si el suelo es de material si...

Page 45: ...ico provocado por transmisores de RF fijos se deber a considerar la posibilidad de realizar una prueba electromagn tica in situ Si la intensidad de campo medida en el lugar donde habr de usarse la Mes...

Page 46: ...18 11 Dimensional...

Page 47: ...la fecha de su publicaci n Los perfeccionamientos t cnicos futuros no resultan en ning n derecho de cambios a productos ya existentes Las im genes presentadas en este manual son para fines ilustrativ...

Page 48: ...t sica inflamable con aire O2 u xido Nitroso Tampoco es adecuado para uso en el entorno rico en Ox geno Este equipo se debe operar solamente por m dicos y enfermeros para ex menes cl nicos ambulatorio...

Page 49: ...e y almacenamiento 00 a 45 C No mueva y no guarde el equipo en superficies irregulares Este equipo est contraindicado para cualquier uso distinto de aquel para el que se pretende o para ser operado po...

Page 50: ...ones Verificaci n de las abrazaderas izquierda y derecha ART CULOS OPCIONALES Verificaci n de las articulaciones del soporte de colposcopio Verificaci n de las abrazaderas dobles Verificaci n de los a...

Page 51: ...n a la red el ctrica de energ a con tensi n incorrecta 5 Esta garant a cesar a Por el decurso normal de su plazo de validez b Por alteraciones no autorizadas por Olsen en el producto c Por adulteraci...

Page 52: ...de mantenimiento muy bajo Estas cualidades se alcanzar n a trav s de un equipo competente y dedicado del cual me enorgullezco sobre todos los aspectos dispuesto a llevar a nuestros clientes lo mejor...

Page 53: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 54: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Page 55: ......

Page 56: ...55 48 2106 6000 export3 olsen odo br Registro en el Ministerio de Salud de Brasil 10281300011 Responsable T cnico Ing Cleber da Costa CREA SC S1 116283 5 C d 5400400 Rev 06 22 03 2021...

Reviews: