15
gently. Insert the file into the tip of the adapter, thread the tip and then use the Endo wrench (5) to tighten. To
remove the adapter, use the Endo wrench (5) to loosen it and then carefully unscrew.
6.3.1
- Scaler Adjustments - Operation and Basic Adjustments
To activate the scaler: push the propulsion/activation
pedal (6).
To adjust the vibration power: turn the control knob
(7) until the desired power is reached.
To toggle functions: push the control knob (7) in direction of
the table for "GP" function (General and Perio). Pull the knob gently in the
opposite direction of the table for "E" (Endo) function.
To adjust the water (including for multifunction display): use the control knob (8) on the instrument control
panel (
section 5.4
).
6.3.2
- Scaler Adjustments - Multifunction Display
When removing the scaler from the holder or extending the
rod (Cross Flex), the multifunctional display (9) screen will identify
the scaler, the number of its position on the working table
(1 to 5), the selected function ("PG" or "E") and will display the
vibration power bar.
To adjust the vibration power: use the keys to increase
(11) or decrease (10) the power, which will be indicated on the
display.
To toggle functions: use the key (12) to toggle between "PG"
(General and Perio) and "E" (Endo) functions indicated in the center
of the multifunctional display.
It is not possible to switch the PG/E functions during the operation of the scaler (while
propulsion/activation pedal is activated).
6.3.3
- Scaler Tips
The tips GD1, PD1 and PD3 accompany the scaler. The power and application information for each tip is
available in the table below.
Note:
1 -
Do not use the tip on features other than those specified by the manufacturer.
2 -
Do not exceed the power limit for each tip. That could damage the tip and the scaler.