14
OLCT10N
Manual
The OLCT10N can be calibrated either :
-
Manually
-
Or automatically
The manual calibration have to be made only after each sensor or
detector replacement. A manual calibration has to be followed by
an automatic calibration.
Please find below the different calibration gas concentrations.
Gas
Range
Calibration gas
cocnentration
Oxygen (O2)
0-30% vol
20.9 %
Carbone monoxide (CO)
0-300 ppm
0-1000 ppm
100 ppm
300 ppm
Hydrogen sulfide (H2S)
0-30 ppm
0-100 ppm
25 ppm
25 ppm
Nitrogen monoxide
0-100 ppm
0-300 ppm
50 ppm
50 ppm
Nitrogen dioxide
0-10 ppm
0-30 ppm
10 ppm
10 ppm
Ammonia 0-100
ppm
0-1000 ppm
100 ppm
1000 ppm
Automatic calibration