anything that will prevent them from measuring the ambient environment for which
they are intended.
The detector shall be installed with the cable gland pointing downwards.
If
installed at an angle of more than 45° from the vertical or in a horizontal position
(ceiling-type installation), the detector will not measure correctly and will need to
be re-calibrated.
Figure 2: OLCT 10N dimensions (in millimeters)
Connection
As a reminder, the OLCT 10N can only be connected to the OLDHAM MX 43 or
MX 32v2 controller.
RS485 transmission
The OLCT 10N has 2 cable glands that are used for the input cable and the output
cable to the next module.
The detectors must be connected using MPI-22A cable type, minimum core size is
0.22 mm² (RS485 cable with 2 shielded twisted pairs, nominal impedance of
100 Ohm). This cable carries the RS485 signal on one pair and the power supply
for the modules on the other pair. The +24VDC, 0V, A and B terminals are
respectively connected to the +24VDC, 0V, A and B terminals of the other