52
OLC(T)100
Technical manual
Type of gas
Measurement
range
(ppm)
XP
Version
IS
Version
Temperature
range°C
% RH
NH
3
Ammonia
100
1000
5000
■
■
■
■
■
■
-20 to +40
15 - 90
NO
Nitrogen
monoxide
100
300
1000
■
■
■
■
■
■
-20 to +50
15
- 90
NO
2
Nitrogen
dioxide
10,0
30,0
■
■
-20 to+ 50
15
- 90
O
2
Oxygen
0-30% vol
■
■
-20 to +50
15 - 90
PH
3
Phosphine
1,00
■
-20 to +40
20 - 90
SiH
4
Silane
50,0
■
-20 to +40
20 - 95
SO
2
Sulfur dioxide
10,0
30,0
100
■
■
■
-20 to +50
15
- 90
Specific characteristics (table 2/2)
Type of
gas
Accuracy
(ppm)
Life
(months)
Response
time
T
50
/ T
90
(s)
Storage
conditions
Warm-up
time max
(h)
AsH
3
+/- 0,05
18
30/120
(1)
1
Cl
2
+/- 0,4
24
10/60
(1)
1
ClO
2
+/- 0,3
24
20/120
(1)
1
CO
+/- 3
(range 0-100)
36
15/40
(1)
1
COCl
2
+/- 0,05
12
60/180
(2)
1
ETO
+/- 1
36
50/240
(1)
24
H
2
+/-5 %
24
30/50
(1)
1
H
2
S
+/- 1,5
(range 0-30)
36
15/30
(1)
1
HCl
+/- 0,4
(range 0-10)
24
30/150
(1)
24
NH
3
+/- 5
+/- 20
+/-150 or 10%
24
25/70
20/60
60/180
(1)
1
NO
+/- 2
(range 100)
36
10/30
(1)
1