6 - Maintenance
35
Possible errors
The table below summarizes the various possible detector errors:
OLC100 explosimeter
Observed fault
Possible cause
Action
Zero setting not
possible
Sensor
Change sensor
Cable
Check cable
Main unit detector module
Check module
Sensitivity
adjustment not
possible
Sensor
Connector cable
Inappropriate standard gas
Change sensor
Check cable
Check standard gas
concentration
High concentration
of explosive gases
Déréglage
Zero setting
Calibration
OLCT100 Detector
Observed fault
Possible cause
Action
Line current 0 mA
Connector cable
Power supply
PCB
Check cable
Check voltage
Change PCB
0 mA < Line
current < 1mA
Sensor
PCB
Line resistance too high
Power supply
Change sensor
Change PCB
Check cable
Check voltage
Zero setting not
possible
Sensor
PCB
Change sensor
Change PCB
Sensitivity
adjustment not
possible
Sensor
PCB
Change sensor
Change PCB
High concentration
Settings disturbed
Zero setting
Calibration
030