91
9.3.2. Device transfer curves in 0% to 30.0% OXYGEN configuration
The following curve shows the response of the device in terms of value measured, and fault
processing as a function of the input current value supplied by the detector. Where the user connects
a brand of detector other than OLDHAM to the MX52 device, he must check carefully that the
transfer curve is fully compatible with the device input characteristics, to ensure that the data
generated by the detector is correctly interpreted. Equally, the device must supply a suitable power
supply voltage, allowing for cable voltage losses.
9.3.3. Power supply and load resistance characteristics
Maximum current available between terminals 2 and 3: 350 mA at 21 V.
Maximum no-load voltage between terminals 2 and 3: 30 V
Load resistance (outside the IS barrier) between terminals 1 and 2: 47 ohms
N.B.: This data applies only where detectors other than OLDHAM are used. Where different
types are mixed, please contact OLDHAM to establish the feasibility of the combination.
9.4. MARKING
OLDHAM
0080
II 2 (G)
INERIS 04ATEX0064
4
mA
20
mA
Signal supplied by
the detector in mA
Display in %
O2 by volume
Fault
Fault
1.2
mA
0.0
%
-5.3%
30.0%
36.0%
23.2
mA
Summary of Contents for MX 52
Page 4: ...4 ...
Page 44: ...44 ...
Page 56: ...56 ...
Page 57: ...57 7 7 V VI IE EW WS S S SP PE EC CI IF FI IE ED D I IN N T TH HE E M MA AN NU UA AL L ...
Page 58: ...58 ...
Page 59: ...59 ...
Page 60: ...60 ...
Page 61: ...61 ...
Page 62: ...62 ...
Page 63: ...63 ...
Page 64: ...64 ...
Page 65: ...65 ...
Page 66: ...66 ...
Page 67: ...67 ...
Page 68: ...68 2 ...
Page 69: ...69 ...
Page 70: ...70 ...
Page 71: ...71 ...
Page 72: ...72 ...
Page 73: ...73 ...
Page 74: ...74 ...
Page 75: ...75 ...
Page 76: ...76 ...
Page 77: ...77 ...
Page 78: ...78 ...
Page 79: ...79 ...
Page 80: ...80 ...
Page 81: ...81 ...
Page 82: ...82 ...
Page 83: ...83 ...
Page 84: ...84 ...
Page 85: ...85 ...
Page 86: ...86 ...
Page 92: ...92 ...
Page 93: ...93 ...
Page 94: ...94 ...
Page 95: ......