- 13 -
1.4.2.
4-20 mA current outputs (Fig. 12)
For each measuring channel, the MX52 unit is equipped with a 4-20 mA output that can be
used to retransmit measurements to a recorder or an external PLC. The maximum resistance
in loop mode is 600 ohms. The earth connections for the 4-20 mA outputs are common and
the unit. The 4-20 mA lines are not galvanically insulated one from the other. The current
output varies according to the measurement and has several states, as follows:
-
On starting up the unit: I < 1 mA
-
With FAULT: I < 1 mA
-
In MAINTENANCE mode: I = 2 mA
-
ZERO MEASUREMENT: I = 4 mA
-
Full scale: I = 20 mA
-
Out of range or “in doubt”: I > 23.2 mA
An example of the connection of a multi-channel recorder is given in Fig. 25.
1.4.3.
RS 232 and RS 485 outputs
RS232 OUTPUT
A computer can be connected on a female sub.D/DB9 type connector located on the back of
the micro board (fig6 repA). The MX52 programming, from outside, will be possible thanks
to this connection.
RS 232 OUTPUT USING
-
Remove the DB9 connector (pmug with an internal strap)
-
Connect a link cable ref.6315831 which will link the monitor to the computer on
the MX52 available female connector DB9 (repA Fig6)
MX52
ORDINATEUR PC
Cordon liaison PC/MX52
( 1 )
( 2 )
-
when the using is stopped : no connect the cable and put the male DB9 "plug"
again.
Female DB9
Male DB9
MX52
PC COMPUTER
Summary of Contents for MX 52
Page 4: ... 4 ...
Page 18: ... 18 ...
Page 22: ... 22 ...
Page 26: ... 26 ...
Page 27: ... 27 ...
Page 28: ... 28 ...
Page 29: ... 29 6 6 V Vi ie ew ws s d de ep pi ic ct te ed d i in n t th he e l le ea af fl le et t ...
Page 30: ... 30 MX52 OVERALL DIMENSIONS FIGURE 1 ...
Page 33: ... 33 B MICRO BOARD FIGURE 6 A ...
Page 47: ......