26
MX 32
Instruction Manual
After initialization
Action or result
Illustration
■
The
CAL
indicator blinks.
Selecting the channel to calibrate
■
Press
(
ENTER
) to confirm.
The
Channel 1
icon blinks indicating
Channel 1
has been selected.
■
To select
Channel 2
, press
(
+
).
The
Channel 2
icon blinks indicating
Channel 2
has been selected.
Zero settings
■
Confirm the channel to calibrate by pressing
(
ENTER
).
- Display of the current value.
X0
= zero settings.
- The
icon is displayed.
- The yellow indicator of the corresponding
channel blinks (the channel relays are
blocked).
■
Make sure that the air is clean, if not, using a test
kit, inject air at the level of the detector at a flow
rate of 60 l/h, then wait for stabilization of the
signal.
If the display is not zero, adjust the zero
measurement by moving the zero potentiometer
corresponding to the appropriate channel (Figure
4, ref. R for
Channel 1
or ref. O for
Channel 2
).
■
Confirm the zero by pressing
(
ENTER
).