![background image](http://html.mh-extra.com/html/oldham/bm25/bm25_user-manual_730578057.webp)
6
–
COM2100
software
57
Alarm relay configuration and logic inputs
■
Set relay output and the logic input by
checking the different boxes (Figure 45).
Reference 1
Alarm relay output settings. In this example
all CO, O2 and LEL alarms will trigger the
relay output.
Reference 3
To set the BM 25/25W modes when logic
input is activated (set to 1).
-
Alarm relay
: internal relay output is
activated
-
Alarm 1
: BM 25/25W will turn in Alarm 1
mode
-
Alarm 2
: BM 25/25W will turn in Alarm 2
mode
-
Remote alarm
: BM 25/25W will turn in
Alarm Transfer mode
Reference 4
Alarm acknowledgment
-
Local acknowledgment
: alarms must be
acknowledged from the gas monitor
keyboard.
-
Remote acknowledgment
: when on a
network (wired or wireless), local alarms
can be acknowledged from another BM
25/25W.
■
Click
OK
to confirm.
Figure 45: alarm output relay and
logic input settings
Summary of Contents for BM25
Page 1: ...BM 25 25W Area Gas Monitor Part Number NPB25EN Version L 2 User Manual ...
Page 8: ...8 BM 25 25W Instruction Manual ...
Page 10: ...10 BM 25 25W Instruction Manual Product Overview Figure 2 Product Overview ...
Page 18: ...18 BM 25 25W Instruction Manual ...
Page 34: ...34 BM 25 25W Instruction Manual ...
Page 70: ...70 BM 25 25W Instruction Manual ...
Page 76: ...76 BM 25 25W Instruction Manual ...
Page 78: ...78 BM 25 25W Instruction Manual ...