6
–
COM2100
software
55
Oldham recommends using pure gases: using mixed gases can alter the
accuracy of gas measurements due to cross interferences between the
sensors.
The calibration gas concentration value shall be between 15% and 100%
of the measurement range.
Date and Time
Settings
■
From
the
Maintenance
menu
(Figure 38), select the option
Update the BM 25/25W date and
time from a PC
.
■
Click
OK
to validate and exit.
Figure 43: the
Time Settings
menu
Monitoring Report
menu
Follow the steps below:
■
From the
Maintenance
menu
(Figure 38), select
’Checking
card’
.
■
Fill in the fields (user's
information for instance)
■
Create the monitoring report
(.ctr file)
■
From the PC (
COM 2100
files)
open the files to edit or print.
Figure 44: the
Monitoring Report
menu