– Special Instructions
71
Chapter 9
| Special Instructions for
use in explosive
atmospheres or
hazardous locations
Information in following paragraphs must be taken into account and followed by the
person responsible for the equipment installation site. Refer to the provisions of
European ATEX Directive 1999/92/EC or to the applicable local legislation, relevant to
improving safety protection and health of workers exposed to the risks of explosive
atmospheres.
The installation and maintenance of the gas monitor should be performed according to
standards EN/IEC 60079-14 and EN/IEC 60079-17.
For intrinsically safe installations and especially for connections to the gas monitor, the
person responsible for the intrinsic safety installation, called the
System designer
, must
establish a system document demonstrating that the entire gas monitor system - Cable -
Body is Intrinsically Safe (see standard IEC 60079-25 for the preparation of this
document).
The BM 25A/AW complies with the following standards:
ATEX
IECEx
Canadian Electrical Code
EN 60079-0:2012/A11:2013 IEC 60079-0 (ed. 6), 2011
C22.2 No. 30
EN 60079-1:2014
IEC 60079-1 (ed. 7), 2014
C22.2 No. 152
EN 60079-11:2012
IEC 60079-11 (ed. 6), 2011
C22.2 No. 157
EN 60079-26:2015
IEC 60079-26 (ed. 3), 2014
CAN/CSA-E79-0
EN 50303:2000
CAN/CSA-E79-1
CAN/CSA-E79-11
Summary of Contents for BM 25A/AW
Page 1: ...BM 25A AW Area Gas Monitor Part Number NP25AEN Version C 4 User Manual ...
Page 10: ...10 BM 25A AW Instruction Manual Product Overview Figure 2 Product Overview ...
Page 18: ...18 BM 25A AW Instruction Manual ...
Page 34: ...34 BM 25A AW Instruction Manual ...
Page 70: ...70 BM 25A AW Instruction Manual ...
Page 75: ...9 Special Instructions 75 CSA Control Drawing for BM 25A AW Connections ...
Page 78: ...78 BM 25A AW Instruction Manual ...