![Ohmeda Ohio Care Plus Operation And Maintenance Manual Download Page 78](http://html1.mh-extra.com/html/ohmeda/ohio-care-plus/ohio-care-plus_operation-and-maintenance-manual_3537110078.webp)
A-8
Appendix
A-8
6600-0239-000 100
This series of bits represents the alarm LEDs. If an LED is illuminated, the corresponding bit is
set to 1 (alarm active). If there is no alarm, the bit is set to 0. Audible alarm status bit is 1 if an
alarm condition exists and if the alarm is not silenced. Audible alarm status bit is 0 if there are
no active alarms, or if the active alarms are silenced.
00000100
Patient Temperature Alarm
Control Temperature Alarm
High Air Temperature Alarm
Air Circulation Alarm
Probe Failure Alarm
System Failure Alarm
Not used
Audible Alarm Status Bit
code
This is the error code that appears in the control temperature display during a system failure
alarm (e.g. E014). If the system is operating normally (no system failure), zeroes replace the
error code (e.g. E000).
ck
This is the two byte ASCII representation of the byte that when added to the sum of all the
ASCII data bytes in the string equals zero. Note that all over flows are dropped and the sum
of the data bytes DOES NOT INCLUDE the <stx>, checksum, <etx>, <cr>, or <lf> characters or
the parity bit of each byte.
<cr>
Carriage return character.
<lf>
Line feed character.
<etx>
End of transmission character (ASCII 3).
Figure A-1
Serial data
D. Nurse Call
Summary of Contents for Ohio Care Plus
Page 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Page 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Page 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Page 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Page 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Page 54: ...5 18 5 18 6600 0239 000 100 Notes...
Page 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Page 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...