2-2
2/Getting Started
2-2
6600-0239-000 100
Mounting the Care Plus on the Elevating or Non-elevating Base
WARNING
Safely mounting or dismounting the Care Plus Incubator or the Elevating Base
rail system requires two people.
1. Verify that O-rings or foam gaskets are installed in the depression around each of
the incubator mounting holes in the incubator platform.
2. Lift the Care Plus Incubator onto the Base. Make sure that the front of the Elevating
Base (foot pedals) faces the same direction as the front of the incubator. Refer to
Figure 2-1.
3. Install the four mounting screws.
WARNING
If mounting hardware is not securely fastened, the incubator could tip off the
stand.
CAUTION
When handling the controller, avoid bumping the fan or the heater. If these items are
knocked out of alignment, the fan can grate against the heater or the base.
4. Install the controller. Lift the levers on the controller and slide it into the incubator as
shown in Figure 2-1. Push the levers down to lock the controller in place.
5. Refer to section 4 in this manual to perform preoperative checkout procedures.
Figure 2-1
Mounting the Care Plus Incubator
Incubator
Platform
Controller
Lever
(Lift to Remove
Controller)
Foot Pedal
Controls
Mounting
Screw
CI.07.049
Summary of Contents for Ohio Care Plus
Page 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Page 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Page 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Page 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Page 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Page 54: ...5 18 5 18 6600 0239 000 100 Notes...
Page 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Page 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...