EN-22
Valor 5000 Series Scales
TABLE 4-2. ERROR CODES.
Err 1 (Overload)
Load exceeds capacity.
Reduce load.
Err 2 (Underload)
Missing pan.
Install the pan and press the zero
button.
Err 9 (Checksum error)
Configuration data
checksum error.
Service required.
Err14 (Zero error)
Zero limit exceeded
during power up.
Remove load from pan.
Err 21 (Checksum error)
Calibration data
checksum error.
Check menu settings and recalibrate.
If error persists, service is required.
CAL E (Calibration error)
Wrong calibration weight
values used.
Calibrate using correct calibration
weights.
4.4 Service Information
If the Troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus
Service Agent. For Service assistance in the United States and Canada, please call Ohaus Corporation toll
free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An Ohaus Product Service Specialist will be
available to help you.
4.5 Accessories
Description
Part
Number
Foot
Switch 71169633