CHAPTER 3 MAINTENANCE PROCEDURES
Valor
™ 1000W Series Service Manual
3-5
Ohaus Corporation www.ohaus.com
TABLE 3-1. MOUNTING BOLT TORQUE SETTINGS
MODEL
TORQUE SETTING
V11PW3
12~13Nm
V11PW6
13~14Nm
V11PW15
14~15Nm
V11PW30
15~16Nm
3.7 Removing/Replacing the Rechargeable Battery
The Valor 1000W has a rechargeable battery.
If it fails to recharge, replace it as follows:
1. Remove the five screws of the bottom
cover, and lift off the cover.
2.
Lift the battery’s front end enough to
access the quick-connect tabs holding
the wire leads.
Note:
Be careful not to short-circuit the
battery leads.
3. Pull the quick-connect tabs free of
the battery posts, and lift out the
battery.
Figure 3-7. Rechargeable battery.
4. Position the replacement battery in the compartment on an angle sufficient to access
the battery posts.
5. Insert the quick-connect tabs on their respective posts, and place the battery fully in
the battery compartment.
6. Position the battery cover, insert the screws and tighten them.
3.10 Replacing the Function Label
The Function Label may need to be replaced. (See Chapter 5 for parts information.)
Use a broad knife to remove the label. Clean the glue residue from the Housing surface.
Then carefully place the new label where the old one was.