OPERATION 

1. Turning Scale On/Off 

To turn scale on, flip toggle switch below right front keyboard. Scale performs a display test, and 
then enters the active weighing mode. To turn the scale off, toggle same switch to off position. 

2. Zero Operation

 

Zero is set automatically at Power On (initial zero). To set it manually press 

ZERO

.   

The scale must be stable and /- 2% of capacity to accept zero operation. 

3. Manual Tare 

Place a container on the weighing pan, and press 

TARE

. The display will show a net value of 

“0”. To clear the Tare value, press 

TARE

 with the pan empty. 

4. Weighing Mode 

Use this mode to measure the weight of a sample. 
To change units, press and hold 

SELECT

; press 

M+ Unit

 until the desired unit icon is displayed. 

5. Accumulate Mode 

Use this mode to store the weight and HI, LO, SUM statistics from a series of samples. 
(See Instruction Manual for details.)

 

6. Check Weigh Mode 

Use this mode to compare samples to a target weight range.  
(See Instruction Manual for details.)

 

7. Setting the Backlight 

Press and hold 

ZERO

 button until 

[BL---x]

 is displayed, where x= 

0

 = backlight is turned off 

1

 = backlight is on during weighing 

2

 = backlight is always on 

Press 

TARE

 to change the setting. Press 

ZERO

 to confirm the setting and return to weighing. 

 
Verify the scale: place a known mass on it and check if the reading is accurate.  
If not, calibrate, using the Calibration Guide. 
 
 
 
 
 
 
 
 
 
 
 
 

*