EN-13
3
000 Series Indicators
3.2 Menu Navigation
3.3 Calibration Menu
Three calibration processes are available: Zero Calibration, Span Calibration and
Linearity Calibration.
Zero Perform
Span Perform
Linearity Perform
Geographic
Adjustment Set 00…
Set 19
… Set 31
End Calibration Exit CALIBRATE menu
NOTES
:
1. Make sure that appropriate calibration masses are available before
beginning calibration.
2. Make sure that the scale base is level and stable during the entire
calibration process.
3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes after
stabilizing to room temperature.
5. To abort calibration, press the
Exit
button anytime during the calibration
process.
6. Make sure to finish Zero Calibration before performing Span Calibration to
ensure weighing accuracy.
TO ENTER THE MENU MODE
Press and hold the Menu button until MENU appears on the display. The first upper level menu appears on the display.
Summary of button navigation functions in menu mode:
--
Yes
Allows entry into the displayed menu.
- Accepts the displayed setting and advances to the next menu item.
--
No
Skips by the displayed menu.
- Rejects the displayed setting or menu item and advances to the next available item.
--
Back
Moves backwards through the upper and middle level menus.
- Backs out of a list of selectable items to the previous middle level menu.
--
Exit
Exits from menu directly to the active weighing mode.
3.3.1 Zero Calibration
Zero Calibration uses one point. The calibration point is established with no load on the scale. Use this
calibration method to adjust for a different pre-load without affecting the span or linearity calibration.
When [CAL] is displayed, press the YES key to accept the Calibration sub-menu selection.
(Pressing the NO key to advance to the next sub-menu, [SEtUP].)
When [ZErO] is displayed, press the YES key to accept the Zero Calibration menu item selection.
The display flashes [0] and the kg led is light.
Press the Yes key to establish the zero point.
Note: The new zero point must be within the range of the normalized weight.
The display shows [--C--] while the zero point is established.
If zero calibration was successful, the scale exits to the next Calibration menu and displays [SPAN]
CAL
ZErO