APPENDIX B SOFTWARE SERVICE TOOL INSTRUCTIONS
Ohaus Corporation www.ohaus.com
B-4
ST5000 Bench Top pH Meter Service Manual
B.4 Configure AppProgramer:
1. Setting the COM Port Configuration in this option you will need to set the Com Port and Baud
Rate. After that click on COM Port Testing to test the communication between computer with
ST5000. Remember both Service Tools and ST5000 should have the same Baud Rate setting.
Figure B-4. COM Port Configuration.
2. Image File Selection is Pre-Fix at file type *.hex. After that you need to browse where you store
the ST5000 *.hex image file.
Figure B-5. Image File Selection.