CHAPTER 3 GETTING STARTED
Ohaus Corporation
www.ohaus.com
3-2
ST5000 Bench Top pH Meter
Service Manual
3. Avoid straining the cables that connect the Main PCB on the bottom housing to the
LCD Touch panel on the top Housing. (Lay the two housings close to each other, so
that the cable is not strained.) (See Figure 3-2)
4. Removing the film cable by tilting up the film cable connector as shown below. (See
Figure 3-3)
5. Replace any parts which are found defective. (Refer to Chapter 4).
Figure 3-2. Connection (marked with yellow
circles) connecting LCD Panel with Main PCBA.
Figure 3-3. Picture on the Left shows the connector at Locked position. Picture on
the Right shows the connector at Un-Locked position.