STARTER 3100M Bench Meter
EN
-
9
3
ST3100M pH Operation
WARNING
Do not operate the equipment in hazardous environments. The
equipment is not explosion protected.
WARNING
When using chemicals and solvents, comply with the instructions of the
chemical producer and the general lab safety rules.
3.1
Setup
3.1.1
Set temperature unit and MTC value
Please note:
If a temperature electrode is used, Automatic Temperature Compensation (
ATC)
and the sample
temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically switches to
Manual Temperature Compensation (
MTC
) mode and
MTC
appears on the screen
.
MTC can be set as follows:
Power on the meter by pressing
.
Press and hold
until the setup icon
appears on the display and the current
temperature unit blinks (
°
C or
°
F).
Press
or
to switch
between
°
C and
°
F .
Press
to confirm your selection.
Then
Continue with MTC temperature setting by using
or
to adjust temperature
compensation accordingly
Press
to confirm the setting
Press
to return to the measurement screen.
The default MTC temperature value setting is 25 °C (77
°
F)
.
Note:
°
C
= 5/9 (
°
F
- 32)
Note:
STARTER 3100M
accepts
NTC 30 kΩ
temperature sensor.